TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
TEMPERATURE CALIBRATION
((
((
(
for technical personnel only)for technical personnel only)
for technical personnel only)for technical personnel only)
for technical personnel only)
The calibration is 2 points at 0.0, 50.0C.
Immerse the probe in a 0C temperature bath.
Press ALT+CALT to enter temperature calibration mode.
The lower LCD displays "0.0 C"; "BUF" and "CAL" tags appear.
When the reading is stable, "CON" symbol starts to blink.
Press ALT+CFM to confirm. The lower LCD displays 50.0C.
Immerse the probe in a 50C temperature bath.
When the reading is stable, "CON" symbol starts to blink.
Press ALT+CFM to confirm and return to normal operation.
TEMPERATURE ADJUSTMENTTEMPERATURE ADJUSTMENT
TEMPERATURE ADJUSTMENTTEMPERATURE ADJUSTMENT
TEMPERATURE ADJUSTMENT
The temperature reading can be manually fine-tuned by following this
procedure:
Press ALT+CALT to enter the temperature
calibration mode.
Press CAL to enter the temperature adjustment mode.
The upper and lower LCD will display the current
temperature reading.
Adjust the temperature reading on
the upper LCD using the arrow keys.
The maximum adjustment is ±1C
around current reading.
Press ALT+CFM to confirm. The meter
returns to measurement mode and dis-
plays the new temperature.
Note: Press ALT+CALT to escape without any changes.
Note: It is possible to enter the temperature adjustment mode only if
the probe is connected.