CALIBRATION CALIBRATION
CALIBRATION CALIBRATION
CALIBRATION
Calibration is 1-point at 100.0% NaCl. Use the HI 7037 calibration
solution (sea water solution) as a 100% NaCl standard solution.
To enter NaCl calibration select the NaCl range and
press the CAL key.
Rinse the probe with some of the calibration solution or
deionized water. Immerse the probe into HI 7037 solu-
tion. The sleeve holes must be completely submerged.
Tap the probe repeatedly to remove any air bubbles that
may be trapped inside the sleeve.
The indications "BUF" and "CAL" are dis-
played. The upper LCD shows the uncalibrated
NaCl reading in percentage. The lower LCD
shows "100".
When the " " symbol stops blinking, the
reading is stable. The "CON" indication starts
blinking on the LCD asking for confirmation.
Press ALT + CFM to confirm the calibra-
tion.
If everything is satisfactory, the meter displays the "Stor Good" mes-
sage and returns to measurement mode.
Note: If the uncalibrated reading is too far from the expected value,
the calibration is not recognized. The "CON" indication does not
appear; the " " and "BUF" symbols blink to signal wrong or
contaminated calibration solution.
Note: The meter uses 1.90%/C compensation factor during calibra-
tion. If the setup item "tc" has been set to a different value,
when exiting the calibration mode the value displayed on the
LCD could be different from the nominal calibration value.
III In order to minimize any EMC interference, use plastic beakers.
IV The meter uses 1.90%/C compensation factor during calibra-
tion. If the setup item "tc" has been set to a different value,
when exiting the calibration mode the value displayed on the
upper LCD could be different from the nominal buffer value.
V It is possible to set the cell constant value directly without
following the calibration procedure. To set the cell constant
enter the setup mode and select "CEL" (see SETUP for details).