User Guide

8
Bottom view
1. DIN connector for EC probe
2. BNC connector for pH electrode
3. Wiring access ports
SPECIFICATIONS
Conductivity
Range 0.00 to 10.00 mS/cm
Resolution 0.01 mS/cm
Accuracy (@20ºC/68ºF) ±5% Full Scale
Typical EMC deviation ±2% Full Scale
Setpoint Adjustable, from 0.50 to 10.00 mS/cm
Temp.Compensation Automatic from 0 to 50°C
Calibration Manual, through slope (80-120%) trimmer
on the front panel
Analog output 0-5V ±5% (0.5V / mS)
Controller output 2A, 220V relay
pH
Range 0.00 to 14.00 pH
Resolution 0.01 pH
Accuracy (@20ºC/68ºF) ±0.02 pH
Typical EMC deviation ±0.1 pH
Setpoint Adjustable, from 0.5 to 14.0 pH
Calibration Manual through offset (±2 pH) and slope
(80 to 120%) trimmers on the front panel