User Guide
33
APPENDIX - A
pH VALUES AT VARIOUS TEMPERATURES.
The temperature directly affects the pH value.
For example, if the buffer’s temperature is 25°C (77°F), calibrate the
meter to read 7.01, 4.01 or 10.01; if the temperature is 20°C, calibrate
the meter to 7.03, 4.00 or 10.06; if the temperature is 50°C, calibrate
the meter to 6.98, 4.06 or 9.82; etc.
Please refer to the following chart for a more accurate pH calibration.
TEMP pH VALUES
°C °F 4.01 6.86 7.01 9.18 10.01
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
32
41
50
59
68
77
86
95
104
113
122
131
140
149
158
4.01
4.00
4.00
4.00
4.00
4.01
4.02
4.03
4.04
4.05
4.06
4.07
4.09
4.11
4.12
6.98
6.95
6.92
6.90
6.88
6.86
6.85
6.84
6.84
6.83
6.83
6.84
6.84
6.85
6.85
7.13
7.10
7.07
7.04
7.03
7.01
7.00
6.99
6.98
6.98
6.98
6.98
6.98
6.99
6.99
9.46
9.39
9.33
9.27
9.22
9.18
9.14
9.10
9.07
9.04
9.01
8.99
8.97
8.95
8.93
10.32
10.24
10.18
10.12
10.06
10.01
9.96
9.92
9.88
9.85
9.82
9.79
9.77
9.76
9.75