User Guide
27
Slope adjustment
• Rinse the electrode and the ground probe
thoroughly with water and immerse the
bottom 4 cm (1.5") in a pH10.01 (HI 7010)
or a pH4.01 (HI 7004) buffer solution.
• Stir the electrode and wait for the reading
to stabilize before adjusting the SLOPE trim-
mer to display the correct pH value at the
measured solution temperature; e.g. 4.01
(or 10.01) at 25°C (77°F). See the “pH
VALUES AT DIFFERENT TEMPERATURES” sec-
tion.
The pH calibration is now complete.
CONDUCTIVITY CALIBRATION
To calibrate the controller, first set it in measurement mode by pressing
the MEA button (the MEA LED lights up).
Make sure the conductivity probe has been properly connected and wired
to the controller and the meter is plugged to the mains.
The calibration should be performed at a temperature similar to that of
the liquid to be monitored.
Slope adjustment
• Pour in a beaker a solution of known conductivity value, which should
be close to the sample stream to be monitored. For example, if
measurements are in the 1.2 to 2.5 EC range, choose HI 7031 (1.41
mS/cm @25°C). Similarly, if measurements are in the 4 to 6 mS/cm
range utilize HI 7031 or HI 7039 (5.00 mS/cm @25°C).
• Immerse the probe in the beaker ensuring that the holes on the probe
sleeve are completely covered.