Conductivity probe connection
• Attach the conductivity probe (HI7632/D or HI3003/D) to the DIN
socket located on the bottom of the controller. Align the guide on the
connector with the socket, push-in the connector and tighten the
retainer ring.
• The probe is provided with a built-in temperature sensor, which allows
automatic temperature compensation of the readings.
Level inputs connections
All level sensors must have a contact type output.
If the water level is above the sensor position, the corresponding contact
is closed. If the water is below the sensor, the contact is open.
Each sensor will be connected between the INPUT COMMON (pin 9) and
the corresponding input pin (LOW LEVEL, NORM LEVEL, HIGH/ALARM
LEVEL).
• If the matching pin is used, connect
the matching pin terminal to the con-
nector pin 2 located on the right panel.
• Immerse the matching pin near the pH electrode.