User Guide

24
MEMORY ORGANIZATION
HI 98801 - HI 98804 only
Logged data are stored in the internal EEPROM and are retained even
if batteries and external power are disconnected.
MEMORY CAPACITY
14000 data samples divided into 16 lots (lots 01 to 16)
9999 data samples maximum for a single channel
300 data samples for the Log on demand (lot 00).
TIMED LOG (lots 01 to 16)
Each time a new logging period starts, it automatically starts from the
next available lot. If the last lot was the 16th, the new logging period
restarts from lot 01 overwriting previously logged data.
When Timed logging memory is full, the meter overwrites the oldest
lot data progressively reducing the old lots. In this case the starting
time, date and the dimension of the old lot are updated.
Note: The oldest lot data are erased without any warning message.
Note: Timed logging memory can be entirely erased through the
setup code 06.
If the meter is powered only by the external power supply and there is
a temporary power black out during logging, when power returns, the
logging continues normally if no samples have been lost, otherwise
the current lot is ended and a new lot starts. If printer is enabled, the
"...Stop..." message will be printed. In any case, during scrolling the
former lot will be preceded by the "Interrupted Lot" message and the
latter by "Continuation Lot" to indicate the interruption.
LOG ON DEMAND (Lot 00)
When Log on demand data area is full the meter shows the "FULL"
message to warn the user that the data are not stored in memory.
Erase the memory area through setup code 05 to continue logging
data on demand.