TO RESET PRINTING SAMPLE NUMBER
(HI98701 and HI98704 only)
This feature resets the print on demand sample number to 001.
Select the code 05
A scrolling message will be displayed.
Press the ALT and CFM keys to confirm reset or ALT and FNC to
escape.
GLP DATA
Cod. 10 - viewing GLP data
Select the code 10
A message will scroll twice across LCD.
The LCD will then display the instrument identification (ID) code.
Press to scan remaining data, in the following order:
channel 1 last calibration date (DD.MM)
channel 1 last calibration year
Note: Data can be viewed in reverse order pressing the key.
Press ALT and FNC to return to function selection mode.
Code 20 - setting the identification (ID) code
When using several identical meters it may be useful to uniquely
identify them by assigning an ID code to each meter.
Select code 20. A message will scroll across LCD.
Enter a 4-digit value using the arrow keys.
Press ALT and CFM to confirm the value.
TESTING BATTERY LEVEL
Select code 41. The message "Battery test" will scroll across LCD.
If the meter is connected to an external power adapter, the LCD
will display "LINE" otherwise it will display bAtt on the upper
display, and the remaining percentage of battery charge (100%
means fully charged battery and 0% cor-
responds to the minimum battery voltage
that allows the meter to operate).
SELECTING THE THERMOCOUPLE TYPE
Three types of thermocouple can be connected to the meter:K, J and T.
Select code 70.
The types are indicated with "dEF" for K-type, "J" for J-type and
"t" for T-type. Choose the desired type using the arrow keys.
Press ALT and CFM to confirm the value.