ElTh ® Pocket-Sized Thermometer
OPERATING: • Remove the protective cap and connect the probe. • Turn the ElTh on by pressing the ON/OFF switch located on the top. • Dip the ElTh probe into solution up to the maximum immersion level. • Stir gently and wait for a few seconds until the display stabilizes. • Read the temperature value of the solution in °C or °F (according to the model) shown on the display. • When not in use, switch ElTh off. • Do not disconnect the probe unless necessary.
rear of the tester using a small screwdriver. Calibration is now complete. BATTERY REPLACEMENT: When the ElTh cannot be switched on or the display fades, pull out the battery compartment and change the four 1.4V batteries, paying attention to their polarity. Batteries should only be replaced in a safe area using the battery types specified in this instruction manual. RECOMMENDATIONS FOR USERS: Before using this product, make sure that it is entirely suitable for the environment in which it is used.
ElTh ® Pocket-Sized Thermometer SPECIFICATIONS: RANGE ElTh °C ElTh °F -50 to 170 °C -58 to 338 °F RESOLUTION ElTh °C ElTh °F 1 °C 1 °F ACCURACY (@20°C/68°F) ElTh °C ±1 °C ElTh °F ±1 °F (-58 to 266 °F) ±2 °F (267 to 338 °F) TYPICAL EMC DEVIATION ElTh °C ±1 °C ElTh °F ±2 °F ADJUSTMENT by trimmer ENVIRONMENT 0 to 50°C 32 to 122°F 95% RH BATTERY LIFE 4 x 1.4V alkaline approx. 700 hours of continuous use DIMENSIONS 142x29x15 mm (5.6x1.2x0.6") WEIGHT 80 g (2.8 oz.