Instruction Manual

27
The answer string for each record contains the number of records if ALL
data is requested - 3 chars
The logged mode (2 chars):
06 - EC range
07 - TDS range
08 - NaCl range
Reading status (1 char): R, O, U
Calculated reading, with sign and decimal point
(7 chars)
Temperature reading, with sign and two decimal
points (8 chars)
EC reading status (1 char): R, O, U - for TDS,
NaCl range
Unit of the reading - for TDS, NaCl range (1 char)
The EC reading, with sign and decimal point - for
TDS, NaCl range (7 chars)
The logged time, yymmddhhmmss (12 chars)
The calibration offset, with sign and decimal point
(7 chars)
The calibration slope, with sign and decimal point
(7 chars) (Kcell for EC and TDS range and the slope
for NaCl).
Temperature probe presence (1 char)
Temperature compensation coefficient (5 chars)
Temperature reference value (5 chars)
Temperature compensation mode (1 char)
Notes: “Err8” is sent if the instrument is not in measurement mode.
“Err6” is sent if the requested range is not available.
“Err4” is sent if the requested set parameter is not available.
“Err3” is sent if the Log on demand is empty.
“Err9” is sent if the battery power is less than 30%.
Invalid commands will be ignored.