Reading status (1 char): R - in range, O - over
range, U - under range.
The reading (corresponding to the selected range) -
7 ASCII chars, including sign and decimal point.
Temperature reading - 8 ASCII chars, with sign and
two decimal points, always in ºC.
MDR Requests the instrument model name and firmware code
(16 ASCII chars).
GLP Requests the calibration data record.
The answer string contains:
GLP status (1 char): represents a 4 bit hexadecimal
encoding.
0x04 - EC calibration available
0x08 - NaCl calibration available
EC calibration data (if available), which contains:
the number of calibrated satandard (1 char)
the offset factor, with sign and decimal point
(7 chars)
the cell constant, with sign and decimal point
(7 chars)
the calibration time, yymmddhhmmss (12 chars)
standards information
standard value, with sign and decimal point
(7 chars).
standard unit (2 chars; 00-µS; 01-mS)
Reference Temperature with and decimal point
(5 chars)
Temperature Compensation mode (2 chars)
00 - no temperature compensation
01 - automatic temperature compensation
02 - manual temperature compensation
TC coeficient with sign and decimal point (5 chars)
Na Cl Calibration data
the number of calibrated standards (1 char)
salinity coeficient, with sign and decimal point
(7 chars)
Cell constant, with sign and decimal point
(7 chars)