Instruction Manual HI 9835 Autoranging Portable Waterproof Microprocessor EC/TDS/NaCl/°C Meter Manufacturers since 1978 This Instrument is in Compliance with the CE Directives
Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the instrument. This manual will provide you with the necessary information for a correct use of the instrument, as well as a precise idea of its versatility. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. These instruments are in compliance with directives EN 50081-1, EN 50082-1 and EN 61010-1. TABLE OF CONTENTS PRELIMINARY EXAMINATION .......
SPECIFICATIONS FUNCTIONAL DESCRIPTION Range EC 0.00 to 29.99 µS/cm (Autoranging) 30.0 to 299.9 µS/cm 300 to 2999 µS/cm 3.00 to 29.99 mS/cm 30.0 to 200.0 mS/cm Up to 500.0 mS/cm actual(*) Conductivity TDS 0.00 to 14.99 ppm (Autoranging) 15.0 to 149.9 ppm 150 to 1499 ppm 1.50 to 14.99 g/L 15.0 to 100.0 g/L Up to 400.0 g/L actual(*) TDS (with 0.80 factor) NaCl 0.0 to 400.0 % Temp. 0.0 to 60.
EC Calibration NaCl Calibration Temperature Calibration Temperature Compensation Temperature Coefficient TDS Factor Probe Auto Off Power supply Casing Environment Dimensions Weight 1 point with 6 memorized buffers 84,1413,5000,12880,80000,111800 µS/cm 1 point with HI 7037 buffer (optional) 2 points at 0 and 50 C (plus ±1 C adjustment) Automatic or Manual from 0 to 60 C (can be disabled to measure actual conductivity) 0.00 to 6.00 %/ C (for EC and TDS only) Default value is 1.90%/ C 0.40 to 0.
AUTORANGING EC/TDS CALIBRATION The EC and TDS scales are autoranging. The meter automatically sets the scale with the highest possible resolution. By pressing ALT+FIXED, the autoranging feature is disabled and the current range is frozen on the LCD. "F1" symbol blinks on the LCD. To restore the autoranging option press ALT+FIXED again. Note: Autoranging is automatically restored if the RANGE key is pressed, if the setup or calibration modes are entered and if the meter is turned off and back on again.
III IV V In order to minimize any EMC interference, use plastic beakers. The meter uses 1.90%/ C compensation factor during calibration. If the setup item "tc" has been set to a different value, when exiting the calibration mode the value displayed on the upper LCD could be different from the nominal buffer value. It is possible to set the cell constant value directly without following the calibration procedure. To set the cell constant enter the setup mode and select "CEL" (see SETUP for details).
CONDUCTIVITY VERSUS TEMPERATURE CHART SETUP The conductivity of an aqueous solution is the measure of its ability to carry an electrical current by means of ionic motion. The conductivity invariably increases with increasing temperature. It is affected by the type and number of ions in the solution and by the viscosity of the solution itself. Both parameters are temperature dependent.
Item tc tcE tdS CEL Aof vEr Chr Description Temp. compensation coeff. Temp. compensation mode TDS factor Cell constant (K) Auto-Off enable Firmware release Battery level test Valid values Default 0.00 to 6.00 %/ C 1.90 Atc, Mtc, notc Atc 0.40 to 0.80 0.50 0.500 to 1.700 1.000 On, OFF On Notes I Once enabled, the Auto Off time is fixed at 5 minutes. II When the battery level test is selected (Chr), LCD will display the remaining percentage of battery charge.
ACCESSORIES CONDUCTIVITY BUFFER SOLUTIONS HI 70030P HI 7030L HI 7030M HI 70031P HI 7031L HI 7031M HI 70033P HI 7033L HI 7033M HI 7034L HI 7034M HI 7035L HI 7035M HI 70039P HI 7039L HI 7039M HI 7037L 12880 µS/cm (µmho/cm), 20mL sachets (25 pcs.) 12880 µS/cm (µmho/cm), 460mL bottle 12880 µS/cm (µmho/cm), 230mL bottle 1413 µS/cm (µmho/cm), 20mL sachets (25 pcs.) 1413 µS/cm (µmho/cm), 460mL bottle 1413 µS/cm (µmho/cm), 230mL bottle 84 µS/cm (µmho/cm), 20mL sachets (25 pcs.
WARRANTY CE DECLARATION OF CONFORMITY All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The electrodes and the probes are warranted for a period of six months. This warranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered.
HANNA LITERATURE Hanna publishes a wide range of catalogs and handbooks for an equally wide range of applications. The reference literature currently covers areas such as: • • • • • • • Water Treatment Process Swimming Pools Agriculture Food Laboratory Thermometry and many others. New reference material is constantly being added to the library. For these and others catalogs, handbooks and leaflets, contact your dealer or the Hanna Customer Service Center nearest to you.