User Guide
Dear Customer,
Thank you for choosing a Hanna product.
Please read this instruction manual carefully before using the in-
strument.
This manual will provide you with the necessary information for a
correct use of the instrument, as well as a precise idea of its
versatility.
If you need additional technical information, do not hesitate to
e-mail us at tech@hannainst.com.
These instruments are in compliance with directives.
TAKING MEASUREMENTS
The meter is factory calibrated for ppm and temeprature.
Press the ON/OFF key to turn
the meter on.
Immerse the probe into the so-
lution to be tested. The
minimum water level must be
approx. of 4½ cm, while the
maximum level has not to ex-
ceed the indicated hole. If
possible, tap the probe to re-
move any air bubbles that may
be trapped inside.
Allow for the reading to stabi-
lize. The LCD displays the
ppm reading.
Press the RANGE key to view
the temperature reading.
Press RANGE again to return
to ppm reading.
If the ppm reading is out of range, the “
---
” indication
is displayed. If the temperature reading is out of range,
the “°C” symbol blinks.
HOLD FUNCTION
To freeze the current measure-
ment on the LCD, press and
hold the RANGE/HOLD key for
about 3 seconds.
When the hold function is active, the LCD displays
alternatively the “HLd” indication and the frozen read-
ing.
Press RANGE to switch from ppm to temperature read-
ing.
Press the ON/OFF key to exit the hold mode and
return to normal operation.
BATTERY REPLACEMENT
To prolong the battery life, the meter switches itself off
after 5 minutes of non-use.
Variations in readings or fading display can indicate
weak batteries.
If the batteries are too weak to support proper opera-
tions, the LCD displays the message “Eb” and then the
meter switches itself off.
Battery replacement must only take place in a non-
hazardous area using AAA-type 1.5V, alkaline batteries.
In order to replace run down
batteries, simply remove the
two screws on the bottom
cover and replace the two 1.5V
AAA batteries with new ones,
paying attention to the correct
polarity.
New batteries allow approx. 1000 hours of continuous
use (with 1175mA/h batteries).