User Guide

For better accuracy, frequent calibration of the instru-
ment is recommended. In addition, the instrument
must be recalibrated whenever:
a) After testing aggressive chemicals.
b) Where high accuracy is required.
c) At least once a month.
To change the EC/TDS conversion factor (CONV)To change the EC/TDS conversion factor (CONV)
To change the EC/TDS conversion factor (CONV)To change the EC/TDS conversion factor (CONV)
To change the EC/TDS conversion factor (CONV)
and the temperature compensation coefficient and the temperature compensation coefficient
and the temperature compensation coefficient and the temperature compensation coefficient
and the temperature compensation coefficient
ββ
ββ
β
(BETA)(BETA)
(BETA)(BETA)
(BETA)
• From measurement mode, press and hold the
MODE button until
TEMP and the current tempera-
ture unit are displayed on the lower LCD. Eg.
TEMP
°C.
• Press the MODE button again to show the current
conversion factor. Eg.
0.50 CONV.
• Press the SET/HOLD button to change the conver-
sion factor.
• Press the MODE button to show the current tem-
perature compensation coefficient
β. Eg. 2.1 BETA.
• Press the SET/HOLD button to change the tempera-
ture compensation coefficient β.
• Press the MODE button to return to normal mea-
surement mode.
Calibration procedureCalibration procedure
Calibration procedureCalibration procedure
Calibration procedure
• From measurement mode, press and hold the
MODE button until
CAL is displayed on the lower
LCD.
• Release the button and fill the cell cup with the
proper calibration solution: HI7031 (1413 µS/cm) for
HI 98321 and HI7030 (12.88 mS/cm) for HI 98322.
• Once the calibration has been automatically per-
formed, the LCD will display
OK for 1 second and the
meter will return to normal measurement mode.
• Since there is a known relationship between EC and
TDS readings, it is not necessary to calibrate the meter
in TDS. If the EC/TDS conversion factor is either 0.5
or 0.7, the meter will allow a direct calibration in
ppm by using the Hanna calibration solutions listed
below.
The
CAL tag on the LCD means that the meter is
calibrated.
To reset to the default calibrationTo reset to the default calibration
To reset to the default calibrationTo reset to the default calibration
To reset to the default calibration
To clear a previous calibration, press the MODE
button after entering the calibration mode. The lower
LCD will display
ESC for 1 second and the meter will
return to normal measurement mode. The
CAL symbol
on the LCD will disappear. The meter will be reset to
the default calibration.
CALIBRATIONCALIBRATION
CALIBRATIONCALIBRATION
CALIBRATION
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
The meter displays the remaining battery percentage
every time it is switched on. When the battery level is
below 5%, the
- +
symbol on the bottom left of the LCD
lights up to indicate a low battery condition. The batter-
ies must be immediately replaced. If however the battery
level is so low as to cause erroneous reading, the Battery
Error Prevention System (BEPS) will automatically turn
the meter off.
To change the batteries, remove the 4 screws located on
the bottom of the meter.
Remove the screw and the washer in the battery com-
partment and carefully replace the 4 batteries located in
the compartment while paying attention to their polarity.
Replace the washer and tighten the screw in the com-
partment; replace the bottom, making sure that the
gasket is properly seated in place, and tighten the screws.
Note: Batteries should only be replaced in a nonhaz-
ardous area using the battery type specified in this in-
struction manual.
ToTo
ToTo
To
turnturn
turnturn
turn
thethe
thethe
the
metermeter
metermeter
meter
onon
onon
on
andand
andand
and
toto
toto
to
checkcheck
checkcheck
check
batterybattery
batterybattery
battery
statusstatus
statusstatus
status
Press and hold the
MODE button for 2-3 seconds. All
the used segments on the LCD will be visible for a few
seconds, followed by a percent indication of the
remaining battery life. Eg.
% 100 BATT.
Taking measurementsTaking measurements
Taking measurementsTaking measurements
Taking measurements
Rinse and fill the cell cup
with the solution to be tested.
Select either EC or TDS
mode with the SET/HOLD
button.
The measurements should be
taken when the stability sym-
bol ¹ on the top left of the
LCD disappears.
The EC (or TDS) value auto-
matically compensated for
temperature is shown on the
primary LCD while the second-
ary LCD shows the temperature
of the sample.
ToTo
ToTo
To
change the temperature unitchange the temperature unit
change the temperature unitchange the temperature unit
change the temperature unit
To change the temperature unit (from °C to °F), from
measurement mode press and hold the MODE button
until
TEMP and the current temperature unit are
displayed on the lower LCD. Eg.
TEMP °C.
Use the SET/HOLD button to change the temperature
unit, and then press MODE button twice to return to
normal measurement mode.
ToTo
ToTo
To
freezefreeze
freezefreeze
freeze
thethe
thethe
the
displaydisplay
displaydisplay
display
Press the SET/HOLD button for 2-3 seconds until
HOLD appears on the secondary display.
Press either button to return to normal mode.
ToTo
ToTo
To
turnturn
turnturn
turn
thethe
thethe
the
metermeter
metermeter
meter
offoff
offoff
off
Press the
MODE button while in normal measurement
mode.
OFF will appear on the lower part of the
display. Release the button.
NotesNotes
NotesNotes
Notes:
• Before taking any measurement make sure the
meter has been calibrated.
• If measurements are taken in different samples
successively, rinse the cell cup thoroughly to eliminate
cross-contamination; and after cleaning, rinse it with
some of the sample to be measured.
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
ACCESSORIESACCESSORIES
ACCESSORIESACCESSORIES
ACCESSORIES
HI 70030P 12.88 mS/cm @25ºC calibration solu-
tion, 20 mL sachet (25 pcs)
HI 70031P 1413 µS/cm @25ºC calibration solu-
tion, 20 mL sachet (25 pcs)
HI 70032P 1382 ppm @25ºC calibration solution,
20 mL sachet (25 pcs)
HI 70038P 6.44 ppt @25ºC calibration solution, 20
mL sachet (25 pcs)
HI 70442P 1500 ppm @25ºC calibration solution,
20 mL sachet (25 pcs)