User Guide

SPECIFICATIONS:
RANGE 0.0  99.9 µS/cm
RESOLUTION 0.1 µS/cm
ACCURACY (@20°C/68°F) ±2% f.s.
TYPICAL EMC DEVIATION ±2% f.s.
CALIBRATION Manual, at 1 point through trimmer
CAL. SOLUTION HI70033 (84.0 µS/cm)
TEMPERATURE 0 to 50C (32 to 122ºF)
COMPENSATION with ß=2%/C typical
ENVIRONMENT 0 to 50C (32 to 122F);
95% RH not condensing
BATTERY TYPE 4 x 1.5V alkaline (included)
BATTERY LIFE approx. 250 hours
of continuous use
DIMENSIONS 175 x 41 x 23 mm
(7.9 x 1.8 x 1")
WEIGHT 78 g (2.7 oz.)
ISTPWTR1 07/01
ELECTRODE REPLACEMENT:
The electrode is built-in inside the plastic casing and
can be easily replaced in the following way:
Slide off the battery cover.
Remove the screw on the back of the
PWTPWT
PWTPWT
PWT located
below the battery compartment and slide off the
whole internal part of the tester.
Replace the built-in electrode with a new
HI73308.
Before reinserting the internal circuit into the
new plastic casing, make sure the O-rings (2)
of the temperature sensor (3) are properly
seated (see drawing).
To avoid bending the connections to the con-
ductivity sensors (1), do not apply excessive
force while inserting the circuit. In case of any
problem, reslide and reinsert it.
Internal circuit:
ACCESSORIES:
HI 70033P 84 µS/cm calibration kit
(25 x 20 mL sachets)
HI 7033M 84 µS/cm calibration solution
(230 mL bottle)
HI 73308 Probe for PWT
HI 731326 Calibration screwdriver (20 pcs)
Fasten the screw on the back, below the
battery compartment.
Reinsert the battery compartment cover.
Recalibrate the
PWTPWT
PWTPWT
PWT before performing new
measurements.
PWT
HI98308
Pure Water Tester
with Automatic Temperature
Compensation