User Guide
SPECIFICATIONS:
RANGE 0.0 99.9 µS/cm
RESOLUTION 0.1 µS/cm
ACCURACY (@20°C/68°F) ±2% f.s.
TYPICAL EMC DEVIATION ±2% f.s.
CALIBRATION Manual, at 1 point through trimmer
CAL. SOLUTION HI70033 (84.0 µS/cm)
TEMPERATURE 0 to 50C (32 to 122ºF)
COMPENSATION with ß=2%/C typical
ENVIRONMENT 0 to 50C (32 to 122F);
95% RH not condensing
BATTERY TYPE 4 x 1.5V alkaline (included)
BATTERY LIFE approx. 250 hours
of continuous use
DIMENSIONS 175 x 41 x 23 mm
(7.9 x 1.8 x 1")
WEIGHT 78 g (2.7 oz.)
ISTPWTR1 07/01
ELECTRODE REPLACEMENT:
The electrode is built-in inside the plastic casing and
can be easily replaced in the following way:
• Slide off the battery cover.
• Remove the screw on the back of the
PWTPWT
PWTPWT
PWT located
below the battery compartment and slide off the
whole internal part of the tester.
• Replace the built-in electrode with a new
HI73308.
• Before reinserting the internal circuit into the
new plastic casing, make sure the O-rings (2)
of the temperature sensor (3) are properly
seated (see drawing).
To avoid bending the connections to the con-
ductivity sensors (1), do not apply excessive
force while inserting the circuit. In case of any
problem, reslide and reinsert it.
Internal circuit:
ACCESSORIES:
HI 70033P 84 µS/cm calibration kit
(25 x 20 mL sachets)
HI 7033M 84 µS/cm calibration solution
(230 mL bottle)
HI 73308 Probe for PWT
HI 731326 Calibration screwdriver (20 pcs)
• Fasten the screw on the back, below the
battery compartment.
• Reinsert the battery compartment cover.
• Recalibrate the
PWTPWT
PWTPWT
PWT before performing new
measurements.
PWT
HI98308
Pure Water Tester
with Automatic Temperature
Compensation