User Guide

1110
cod. 40: Auto OFF
cod. 41: Battery test
cod. 50: Baud rate
cod. 60: Release code
cod. 70: Celsius or Fahrenheit
cod. 99: Pass Code
Some of the most important functions are explained below in a step
by step sequence.
TO SCAN LOGGED DATA
COD. 00 - Lot data Printing / Scrolling
Select the code 00.
The message "Lot data Printing" will scroll twice on LCD.
The upper LCD will then display L 00
with the 00 blinking.
Set the desired lot by the arrow keys. L00 is the lot of data of the
"log on demand" and L01 to L16 are the lots of the "timed log".
Press the ALT and CFM keys to confirm the lot number.
If the lot doesn't contain data, the "no data" message will scroll
across the LCD twice and the meter will return to setup mode.
If the lot contains one or more data the
LCD will display the sample number in its
upper part and Sn in the lower part.
Note: In the L00 lot (log on demand) the sample number will be
displayed with 3 digits (001).
Select the sample number to scan by the arrow keys.
Printing logged data
Press ALT and CFM to print logged data.
If the selected sample number is invalid (equal to 0 or bigger
than the number of samples), the "Err" message will be displayed
for a few seconds.
If the sample number is correct, the samples starting from the
selected one to the last sample of the lot will be printed. To stop
printing before the last sample is reached, press and hold the ALT
and PAPER keys until the printer stops.
During printout the LCD will display the sample number being
Code Valid values Default
10 Show pH GLP data
11 Show ORP GLP data (HI98240 only)
12 Print GLP data
13 Calibration alarm time-out 0099 days 00
00 means option disabled
14 Instrument ID code 00009999 0000
30 Current time
1
hh:mm 00:00
31 Current day
1
dd 01
32 Current month
1
MM 01
33 Current year
1
YYYY 1998
40 Auto-Off/Power down time-out 5,10,15,30,45,60 5
41 Battery level test
50 RS232 baud rate 1200, 2400, 4800 4800
60 Firmware version
70 Celsius/Fahrenheit selection C ; F C
99 Password
2
00009999 0000
Note: If a wrong code is entered the "Err" message is displayed on
LCD for a few seconds.
1
The meter automatically checks for entered time/date accuracy as follows:
0hh23; 0mm59; 01dd28/29/30/31; 1MM12; 1998YYYY2097.
2
To change the password, the correct code must be entered first. If the password
has been forgotten, the password protected features are no longer accessible;
in this case contact your nearest Hanna Service Center.
SETUP MESSAGES LIST
cod. 00: Lot data Printing
cod. 01: Lot table Printing
cod. 02: Printer enable
cod. 03: Log Interval
cod. 05: Press "ALT CFM" to delete Lot 00 or "ALT FNC" to escape
cod. 06: Press "ALT CFM" to delete Lot 01-16 or "ALT FNC" to escape
cod. 10: pH GLP
cod. 11: Volt GLP (HI98240 only)
cod. 12: GLP data printing
cod. 13: Calibration alarm time-out
cod. 14: Instrument ID Code
cod. 30: Hour - Minute
cod. 31: Day
cod. 32: Month
cod. 33: Year