User Guide

SPECIFICATIONS:
RANGE 0.00 to 1.00 pNaCl
(58.4 to 5.84 g/L NaCl)
RESOLUTION 0.01 pNaCl
ACCURACY (@20°C/68°F) ±0.02 pNaCl
TYPICAL EMC DEVIATION ±0.04 pNaCl
ENVIRONMENT 0 to 50C (32 to 122F); 95% RH
BATTERY TYPE 4 x 1.5V alkaline (included)
BATTERY LIFE approx. 500 hours of continuous use
DIMENSIONS 175x41x23 mm (7.9x1.8x1")
WEIGHT 78 g (2.7 oz.)
IS98203R2 05/01
ACCESSORIES:
HI 7081L 30 g/L NaCl solution (460 mL bottle)
HI 7081M 30 g/L NaCl solution (230 mL bottle)
HI 7061M Electrode cleaning solution
(230 mL bottle)
HI 73202 Spare electrode
HI 731326 Calibration screwdriver (20 pcs)
Remove the electrode and replace
it with a new HI73202.
Fasten first the screw on the front
(1) to attach the electrode to the
circuit. Insert the two small wires
into their sockets and fasten the
two screws (2 & 3).
The new electrode is supplied with four O-rings
(4). Make sure they are placed properly before
reinserting the circuit into the plastic casing.
Reinsert the internal circuit complete with the
new electrode into the plastic casing.
Fasten the screw on the back, below the
battery compartment.
Reinsert the battery compartment cover.
Recalibrate the
SALINTESTSALINTEST
SALINTESTSALINTEST
SALINTEST before performing
new measurements.
CALIBRATION:
Immerse the tester in
HI7081 (30 g/L NaCl) cali-
bration solution, without ex-
ceeding the maximum im-
mersion level.
Allow the reading to stabi-
lize and with a small screw-
driver turn the calibration
trimmer until the display
shows "0.30".
Calibration is now complete.
MAINTENANCE:
In case of erroneous reading even after an accurate
calibration, the reference junction can be contami-
nated or clogged.
Pull out 2 mm (1/8'') of the
cloth junction to renew the
electrode reference (it is recom-
mended to cut the cloth leav-
ing always at least 2 mm -1/
8'' over the reference compart-
ment) and repeat the calibra-
tion procedure.
The cloth junction can be pulled out approximately
20 times. After that, the electrode can be replaced
(see the Electrode Replacement section).
BATTERY REPLACEMENT:
If display fades or the
SALINTESTSALINTEST
SALINTESTSALINTEST
SALINTEST switches off, the
batteries must be replaced. Slide off the battery
compartment cover and replace all four 1.5V batter-
ies while paying attention to their polarity.
ONE YEAR WARRANTY AND SERVICEABLE:ONE YEAR WARRANTY AND SERVICEABLE:
ONE YEAR WARRANTY AND SERVICEABLE:ONE YEAR WARRANTY AND SERVICEABLE:
ONE YEAR WARRANTY AND SERVICEABLE:
Covered by
one yearone year
one yearone year
one year warranty against defects in work-
manship and materials, this tester is now completely
serviceable. Contact your dealer for further information.
The
SALINTEST SALINTEST
SALINTEST SALINTEST
SALINTEST is in compliance with the CE directives.
Visit our Internet Home Page:
http://www.hannainst.com
ELECTRODE REPLACEMENT:
The electrode can be easily replaced in the following
way:
Slide off the battery cover.
Remove the screw on the back of the
SALINTESTSALINTEST
SALINTESTSALINTEST
SALINTEST
located below the battery compartment and slide
off the whole internal part of the tester.
Remove the fastening screw on the front (1) to
loose the electrode. The internal circuit is con-
nected to the electrode trough two wires (one
for the glass sensor and one for the reference).
Remove the two small screws (2 & 3) which
fasten the two connecting wires to their sock-
ets.
SALINTEST
HI98203
Salinity Concentration
Pocket-sized Meter
Internal circuit: