• When the reading is stable and close to the selected calibration point, the
CFM functional key is displayed.
Press CFM to confirm.
The second expected calibrated point is displayed.
Immerse the probe into the second vessel as close as possible to the reference
thermometer. Allow a few seconds for the probe to stabilize.
Use the ARROW keys to set the calibration point value to that measured
by the reference thermometer.
When the reading is stable and close to the selected calibration point,
CFM functional key is displayed.
Press CFM to confirm. The instrument returns to measurement mode.