Instruction Manual

36
The following table lists the specific range parameters:
Item Description Valid value Default
Calibr. Number of days after Disable, 1 to 7 days Disable
Timeout Calibration warning
(EC, NaCl) is displayed
Out cal Display warning if the Enable/Disable Disable
range check reading is too far from the
(EC range calibration points
only)
Temperature Temperature Probe/Manual Probe
source entry mode
Temperature No TC, Linear No TC
compensation Non Linear
mode
Range select Fix one Automatic, Fix one Automatic
specific range resolution of
EC or resistivity range
Cell constant Manual set of 0.010 to 10.000 1.000
the cell constant
Temperature Set the coeficient 0.00 to 10.00 %/°C 1.00 %/°C
coeficient for linear temperature
compensation
Temperature Reference temperature 15 °C, 20 °C, 25 °C 25 °C
Ref
Temperature unit °C or °F °C
TDS factor 0.40 to 1.00 0.50