ERROR SCREENS
Wrong standard
The calibration cannot be confirmed.
The EC reading is not close to the selected standard. Select another standard
using the ARROW keys.
• After the fourth calibration point is confirmed, immerse the EC electrode into the
fifth standard solution and stir gently.
• If necessary, press the ARROW keys to select a different standard value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected standard, the CFM
functional key is displayed.
• Press CFM to confirm calibration.
• The instrument stores the calibration values and returns to normal mea-
surement mode.
FOUR, THREE or TWO-POINT CALIBRATION
• Proceed as described in “FIVE-POINT CALIBRATION” section.
• Press CAL or ESC after the appropriate accepted calibration point. The instru-
ments will return to measurement mode and will memorize the calibration data.