• buffer value including sign and decimal point (this value is sent only
if there is a 2 point calibration) (6 chars)
• salinity value (3 chars)
• pressure value in mmHg, including sign and decimal point
(11 chars)
• temperature value including sign and decimal point (8 chars)
• calibration time: yymmddhhmmss (12 chars).
PAR Requests the setup parameters setting.
The answer string contains:
• backlight value (1 char)
• contrast value (2 chars)
• instrument ID (4 chars)
• calibration alarm timeout (2 chars)
• Setup information (2 chars): 8 bit hexadecimal encoding:
• 0x01:
- -
- -
- beep is on
• 0x04:
- -
- -
- Celsius / Fahrenheit display (°C if the bit is set)
• 0x10: - manual pressure (1 if activated, 0 otherwise)
• Auto Light-off time (3 chars)
• Auto Power-off time (3 chars)
• Salinity value (3 chars)
• Pressure unit (1 char): 0 - mmHg, 1 - inHg, 2 - atm, 3 - mbar,
4 - psi, 5 - kPa
• BOD configuration values
• sample minimum delta DO, including sign and decimal point
(6 chars)
• sample minimum end DO, including sign and decimal point
(6 chars)
• seed minimum delta DO, including sign and decimal point
(6 chars)
• seed minimum end DO, including sign and decimal point
(6 chars)
•
OUR
configuration values
• minimum time in seconds (4 chars)
• maximum time in seconds (4 chars)
• minimum start DO including sign and decimal point (6 chars)
• minimum end DO including sign and decimal point (6 chars)
• total volume including sign and decimal point (6 chars)
• sample volume including sign and decimal point (6 chars)
•
SOUR
configuration values
• minimum time in seconds (4 chars)