Owner's manual

56
0x10: temperature probe connection
0x20: DO measure unit (0 = %, 1= mg/L)
0x01:
new
GLPGLP
GLPGLP
GLP data available
0x02: new setup parameter
0x04: out of calibration range
0x08: the meter is in autoend mode;
Reading status: R - in range, O - over range, U - under range
The following status flags are sent for all modes except
mode 25
DO reading range flag
temperature reading range flag
pressure reading range flag
OUR/SOUR reading range flag (sent only if in OUR/SOUR
measure mode)
The following status flag is sent only for mode 25
BOD reading range flag
Readings
The following values are sent for all modes except
mode 25
DO reading, including sign and decimal point (8 chars)
temperature, including sign and decimal point (8chars)
pressure value, including sign and decimal point, always
in mmHg (11 chars)
OUR/SOUR reading, including sign and decimal point
(sent only if in OUR/SOUR measure mode) (8 chars)
OUR/SOUR counter (sent only if in OUR/SOUR measure
mode) (4 chars)
The following values are sent only for mode 25
BOD reading, including sign and decimal point (8 chars)
initial DO value, including sign and decimal point [mg/L]
(6 chars)
final DO value, including sign and decimal point [mg/L]
(6 chars)
MDR Requests the instrument model name and firmware code (16 ASCII chars).
GLP Requests the calibration data record.
The answer string contains:
the number of calibrated buffers (1 char)
calibrated buffer unit (0 = %, 1 = mg/L) (1 char)
buffer value including sign and decimal point (6 chars)
calibrated buffer unit (0 = %, 1 = mg/L) (this value is sent only
if there is a 2 point calibration) (1 char)