Owner's manual
56
• 0x10: temperature probe connection
• 0x20: DO measure unit (0 = %, 1= mg/L)
• 0x01:
new
GLPGLP
GLPGLP
GLP data available
• 0x02: new setup parameter
• 0x04: out of calibration range
• 0x08: the meter is in autoend mode;
• Reading status: R - in range, O - over range, U - under range
• The following status flags are sent for all modes except
mode 25
• DO reading range flag
• temperature reading range flag
• pressure reading range flag
• OUR/SOUR reading range flag (sent only if in OUR/SOUR
measure mode)
• The following status flag is sent only for mode 25
• BOD reading range flag
• Readings
• The following values are sent for all modes except
mode 25
• DO reading, including sign and decimal point (8 chars)
• temperature, including sign and decimal point (8chars)
• pressure value, including sign and decimal point, always
in mmHg (11 chars)
• OUR/SOUR reading, including sign and decimal point
(sent only if in OUR/SOUR measure mode) (8 chars)
• OUR/SOUR counter (sent only if in OUR/SOUR measure
mode) (4 chars)
• The following values are sent only for mode 25
• BOD reading, including sign and decimal point (8 chars)
• initial DO value, including sign and decimal point [mg/L]
(6 chars)
• final DO value, including sign and decimal point [mg/L]
(6 chars)
MDR Requests the instrument model name and firmware code (16 ASCII chars).
GLP Requests the calibration data record.
The answer string contains:
• the number of calibrated buffers (1 char)
• calibrated buffer unit (0 = %, 1 = mg/L) (1 char)
• buffer value including sign and decimal point (6 chars)
• calibrated buffer unit (0 = %, 1 = mg/L) (this value is sent only
if there is a 2 point calibration) (1 char)