User Guide

Do not be alarmed if white crystals appear around the
electrode protective cap. This is normal with pH elec-
trodes and they dissolve when rinsed with water.
When not in use, rinse the electrode with water to
minimize contamination and store it with a few drops of
storage (HI 70300) or pH 7 (HI 7007) solution in the
protective cap. Always replace the protective cap after use.
DO NOT USE DISTILLED OR DEIONIZED WATER FOR
STORAGE PURPOSES.
If the electrode has been left dry, soak the tip in a
storage (HI 70300) or pH 7 (HI 7007) solution for at
least one hour to reactivate it.
To minimize clogging and provide longer life for the pH
electrode, it is recommended to clean it monthly. Im-
merse the tip of the electrode in HI 7061 for one hour
and then rinse it with tap water.
TAKING pH MEASUREMENTS
Turn the meter on by connecting the
12 VDC power adapter to the meter
and to the mains.
Remove the protective cap from the
pH electrode and immerse the tip
(4 cm/1½”) of pH electrode in the
sample.
The LCD will show the pH value. Allow the
reading to stabilize and the instrument
will start continuous monitoring.
Note: to prevent damages to the electrode, remove the pH
electrode from the sample before turning the meter off.
ADJUSTING THE SETPOINT
With HI 981402 and HI 981403 you can select your own
setpoint and be alerted with a visual LED alarm when an
abnormal situation arises.
Unscrew and remove the rear panel and gasket seal to
access the MEASURE/SET switch and the setpoint adjust-
ment trimmer.
Move the switch to the left (SET).
For the greatest accuracy, frequent calibration of the instru-
ment is recommended. In addition, the instrument must be
recalibrated whenever:
a) The pH electrode is replaced.
b) After testing aggressive chemicals.
c) Where extreme accuracy is required.
d) At least once a month.
PREPARATION
Pour small quantities of pH
7.0 (HI 7007) and pH 4.0
(HI 7004) solution into two
clean beakers.
For accurate calibration use two beakers for each buffer
solution, the first one for rinsing the tip of the electrode and
the second one for calibration. This way, contamination of
the buffers is minimized.
CALIBRATIONCALIBRATION
CALIBRATIONCALIBRATION
CALIBRATION
HI 7004 HI 7007HI 7004 HI 7007
H
I 7
0
0
4
With a small screwdriver adjust the setpoint trimmer to
display the desired value in the range 3 to 11 pH (HI
981402) or 6 to 9.5 pH (HI 981403).
Make sure the switch is moved back to the
right (MEASURE Mode).
Replace the rear panel and the gasket, ensuring the
unit is properly closed.
In HI 981402, whenever the pH
reading varies by more than ±0.5
pH from the setpoint, the red alarm
LED blinks to warn the user.
In HI 981403 the alarm LED
starts blinking when the pH read-
ing becomes greater than setpoint
and stops blinking only when
the reading becomes 0.5 pH lower
than setpoint.
HI 7007
RINSE
HI 7007
CALIBRATION
pH 4.0 (HI 7004) is recommended for measuring acidic
samples. Use pH 10.0 (HI 7010) if subsequent samples are
alkaline.
CALIBRATION PROCEDURE
Turn the meter on and make sure that
the MEASURE/SET switch is on the MEA-
SURE mode.
Remove the protective cap from the
electrode, rinse and immerse it in a pH
7.0 buffer. Stir gently and then wait a
couple of minutes for the reading to
stabilize.
Note: the electrode should be sub-
merged approximately 4 cm
(1½”) in the solution.
Adjust the right hand trimmer
with the calibration screwdriver
until the LCD shows pH 7.0.
Rinse and immerse the pH electrode in
pH 4.0 (or pH 10.0) buffer and stir
gently.
Wait a couple of minutes and
then adjust the left hand trim-
mer until the LCD shows the
value of the second buffer.
The pH calibration is now complete.
4 cm
(1½
")
CE DECLARATIONCE DECLARATION
CE DECLARATIONCE DECLARATION
CE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
Recommendations for Users
Before using this product, make sure that it is entirely suitable for the environment in
which it is used. Operation of this instrument in residential areas could cause unaccept-
able interferences to radio and TV equipment.
The glass bulb at the end of the electrode is sensitive to electrostatic discharges. Avoid touching
this glass bulb at all times. During operation, ESD wrist straps should be worn to avoid possible
damage to the electrode by electrostatic discharges.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s
EMC performance.
To avoid electrical shock, do not use this instrument when voltages at the measurement surface
exceed 24 VAC or 60 VDC. To avoid damages or burns, do not perform any measurement in
microwave ovens.
ACCESSORIESACCESSORIES
ACCESSORIESACCESSORIES
ACCESSORIES
HI 1286 Double junction, plastic body pH electrode
with 2 m (6.6’) cable and BNC connector
HI 2114P/2 Double junction, plastic body pH electrode
with 2 m (6.6’) cable and BNC connector
HI 70004P pH 4.01 solution, 20 mL sachet (25 pcs)
HI 70007P pH 7.01 solution, 20 mL sachet (25 pcs)
HI 7004L pH 4.01 solution, 460 mL bottle
HI 7007L pH 7.01 solution, 460 mL bottle
HI 7061L Electrode cleaning solution, 460 mL bottle
HI 710005 12 VDC power adapter, US plug
HI 710006 12 VDC power adapter, European plug
HI 710012 12 VDC power adapter, Australian plug
HI 710013 12 VDC power adapter, Southern Africa plug
HI 710014 12 VDC power adapter, UK plug
HI 77400P pH 4 & 7 solutions, 20 mL sachet (5 each)
ISTR9814R1 07/98PRINTED IN PORTUGAL