User Guide

When not in use, rinse the electrode with water to
minimize contamination and store it with a few drops of
storage (HI 70300) or pH7 (HI7007) solution in the
protective cap. Always replace the protective cap after use.
DO NOT USE DISTILLED OR DEIONIZED WATER FOR
STORAGE PURPOSES.
If the electrode has been left dry, soak the tip in a
storage (HI 70300) or pH7 (HI7007) solution for at
least one hour to reactivate it.
To minimize clogging and provide longer life for the pH
electrode, it is recommended to clean it monthly. Im-
merse the tip of the electrode in HI 7061 for one hour
and then rinse it with tap water.
TAKING pH MEASUREMENTS
Turn the meter on by connecting the
12 VDC power adapter to the meter
and to the mains.
In addition to the pH electrode, the
pH is supplied with a
differential input and HI1283 stain-
less steel probe to prevent potential
grounding problems. Remove the pro-
tective cap from the pH electrode and
immerse the tips (4cm/1½) of both
pH electrode and grounding probe in
the sample.
The LCD will show the pH value. Allow the reading to stabilize
and pH will start continuous monitoring.
Note: to prolong the electrode life, after switching the meter
off, remove the electrode from the solution or, alterna-
tively, detach it from the meter.
For the greatest accuracy, frequent calibration of the instru-
ment is recommended. In addition, the instrument must be
recalibrated whenever:
a) The pH electrode is replaced.
b) After testing aggressive chemicals.
c) Extreme accuracy is required.
d) At least once a month.
PREPARATION
Pour small quantities of pH 7.01 (HI7007) and pH 4.01
(HI7004) solution into two clean beakers.
For accurate calibration use two beakers for each buffer
solution, the first one for rinsing the tip of the electrode and
the second one for calibration.
This way, contamination of the buffers is minimized.
pH4.01 (HI7004) is recommended for measuring acidic
samples. Use pH 10.01 (HI7010) if subsequent samples
are alkaline.
CALIBRATION PROCEDURE
Remove the protective cap from the
electrode, rinse and immerse it in a
pH 7.01 buffer with the grounding
probe. Stir gently and then wait a
couple of minutes for the reading to
stabilize.
CALIBRATIONCALIBRATION
CALIBRATIONCALIBRATION
CALIBRATION
HI 7004 HI 7007HI 7004 HI 7007
HI 7004
HI 7007
RINSE
HI 7007
CALIBRATION
Note: the electrode should be sub-
merged approximately 4cm
(1½) in the solution.
Adjust the right hand trimmer
with the calibration screwdriver
until the LCD shows 7.0.
Rinse and immerse the pH electrode
and the grounding probe in pH4.01
(or pH10.01) buffer and stir gently.
Wait a couple of minutes and
then adjust the left hand trim-
mer until the LCD shows the
value of the second buffer.
The pH calibration is now complete.
4 cm
(1½")
CE DECLARATIONCE DECLARATION
CE DECLARATIONCE DECLARATION
CE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
Recommendations for Users
Before using this product, make sure that it is entirely suitable for the environment in
which it is used. Operation of this instrument in residential areas could cause unaccept-
able interferences to radio and TV equipment.
The glass bulb at the end of the electrode is sensitive to electrostatic discharges. Avoid touching
this glass bulb at all times. During operation, ESD wrist straps should be worn to avoid possible
damage to the electrode by electrostatic discharges.
Any variation introduced by the user to the supplied equipment may degrade the instruments
EMC performance.
To avoid electrical shock, do not use this instrument when voltages at the measurement surface
exceed 24 VAC or 60 VDC. To avoid damages or burns, do not perform any measurement in
microwave ovens.
ACCESSORIESACCESSORIES
ACCESSORIESACCESSORIES
ACCESSORIES
HI 1283 * Stainless steel grounding probe with 2 m
(6.6) cable
HI 1286 Double junction, plastic body pH electrode
with 2 m (6.6) cable and BNC connector
HI 70004P pH 4.01 solution, 20 mL sachet (25 pcs)
HI 70007P pH 7.01 solution, 20 mL sachet (25 pcs)
HI 7004L pH 4.01 solution, 460 mL bottle
HI 7007L pH 7.01 solution, 460 mL bottle
HI 7061L Electrode cleaning solution, 460 mL bottle
HI 710005 12 VDC power adapter, US plug
HI 710006 12 VDC power adapter, European plug
HI 710012 12 VDC power adapter, Australian plug
HI 710013 12 VDC power adapter, Southern Africa plug
HI 710014 12 VDC power adapter, UK plug
HI 77400P pH 4 & 7 solutions, 20 mL sachet (5 each)
* To be replaced only by authorized service technicians