INITIAL PREPARATION
Each meter is supplied complete with batteries. Remove the back
cover, unwrap the batteries and install them while paying attention
to the polarity. Alternatively, connect the HI 710005 or HI 710006
voltage adapter to the power adapter plug.
To prepare the instrument for use, connect the pH or ORP (HI98150
only) electrode to the connector located on the top of the instrument.
To switch the meter on, press the ON/OFF key.
The batteries charge status or "LINE" message
(if external power adapter is connected) will be
displayed on the LCD for a few seconds.
The meter is now ready to operate.
To maximize battery life, the meter is automatically switched off after 5
minutes of non-use (this feature is enabled by default and can be
disabled through setup code 20). To reactivate the instrument press
the ON/OFF key.
Before proceeding with pH measurements follow the calibration proce-
dure.
Note: When the use of an alternate function (DEL, FNC, CFM, GLP,
RCL and CALT) is requested, press and hold the ALT key first
and then the second listed key.
Note: to prevent damage to the electrode, remove it from the sample
before turning the meter off.
If the meter is OFF, detach the electrode from the meter before
immersing it in the storage solution.