User Guide

13
These meters are powered by a 9V battery that is located on the rear
of the instrument.
When the battery level is so low that it may cause unreliable
measurements, the meter turns off.
It is recommended to replace the battery immediately.
Battery replacement must only take place in a nonhazardous area
using a 9V alkaline battery.
Unscrew the three screws
on the rear of the meter,
remove the battery com-
partment cover and
replace the 9V battery
with a new one while
paying attention to its
polarity.
Make sure the battery
contacts are tight and
secure before replacing
the cover.
BATTERY REPLACEMENT
4. if the red LED lights up again, calibration is needed (see
"pH Calibration" section).
Note: If it is not possible to calibrate, the probe has to be replaced
with a new one.
Note: For field applications, it is always recommended to keep a
spare probe handy. When anomalies are not resolved with
simple maintenance, change the probe and recalibrate the
meter.