Instruction Manual HI 9813 HI 9813-5 & HI 9813-6 Portable pH/EC/TDS/°C Meters w w w . h a n n a i n s t .
Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the meter. This manual will provide you with the necessary information for correct use of the instrument, as well as a precise idea of its versatility. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. These instruments are in compliance with the directives.
PRELIMINARY EXAMINATION Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipment. If noticeable damage is found, notify your dealer. Each meter is supplied with: • Combination probe with 8-pin DIN connector & 1 m (3.3') cable: HI 1285-0, pH/EC/TDS probe, for HI 9813; HI 1285-5, pH/EC/TDS/°C probe, for HI 9813-5; HI 1285-6, pH/EC/TDS/°C probe with Cal-Check feature, for HI 9813-6 • HI 70007 pH 7.01 sachet, 1 pc • HI 70031 1413 µS/cm (1.
FUNCTIONAL DESCRIPTION HI 9813 AND HI 9813-5 1) 2) 3) 4) 5) 6) 7) 8) 9) 8-pin DIN connector for probe Liquid Crystal Display pH range selection key mS/cm (EC) range selection key EC/TDS calibration knob pH offset calibration knob ppm (TDS) selection key °C (Temperature) selection key, for HI 9813-5 only ON/OFF key 4
FUNCTIONAL DESCRIPTION HI 9813-6 1) 2) 3) 4) 5) 6) 7) 8) 9) 8-pin DIN connector for probe Liquid Crystal Display Green and Red LEDs for Cal-Check mS/cm / ppm (EC/TDS) range selection key EC/TDS calibration knob pH offset calibration knob pH/°C range selection key Check key ON/OFF key 5
SPECIFICATIONS Range pH 0.0 to 14.0 ppm 0 to 1999 mS/cm 0.00 to 4.00 °C 0.0 to 60.0 (HI 9813-5 & HI 9813-6) Resolution pH 0.1 ppm 1 mS/cm 0.01 °C 0.1 (HI 9813-5 & HI 9813-6) Accuracy pH ±0.1 (@20°C/68°F) ppm ±2% f.s. mS/cm ±2% f.s. °C ±0.5 (HI 9813-5 & HI 9813-6) Conversion Factor variable, 0.56 to 0.72 Typical EMC pH ±0.1 Deviation ppm ±2% f.s. mS/cm ±2% f.s. °C ±0.
OPERATIONAL GUIDE INITIAL PREPARATION Each meter is supplied complete with a 9V battery. Remove the battery compartment cover on the back of the meter and install the battery while paying attention to its polarity. Connect the probe to the DIN socket on the top of the meter by aligning the pins with the socket and pushing in the plug. Always remove the probe protective cap before taking any measurements, and make sure the meter has been calibrated. Turn the meter on by pressing the ON/OFF key.
• Select the appropriate measurement range (EC or TDS). • Wait for a couple of minutes for the temperature sensor to reach thermal equilibrium. The display then shows the measurement automatically compensated for temperature with the appropriate following indication: "mS" tag indicates the meter is in EC mode, while no symbol indicates the meter is in TDS mode. TAKING TEMPERATURE MEASUREMENTS (not for HI 9813) • Submerge the tip (4cm/1½") of the probe into the sample to be tested. • Select the °C mode.
pH CALIBRATION For greatest accuracy, frequent calibration of the instrument is recommended. The instrument should be recalibrated for pH: a) Whenever the electrode is replaced. b) At least once a month. c) After testing aggressive chemicals. d) Where extreme accuracy is required. e) Whenever the CAL-CHECK test fails (HI 9813-6 only). PREPARATION Pour small quantities of pH 7.01 (HI 7007) or pH 4.01 (HI 7004) or pH 10.01 (HI 7010) solution into a clean beaker. To obtain accurate readings, use pH 7.
Note: The probe should be submerged approximately 4 cm (1½") into the solution. The ChecktempC thermometer has to be located close to the probe. Note: If turning the knob the needed value can not be reached, clean the probe (see the "Probe Maintenance" section). If also after the probe cleaning the value can not be reached, replace the probe. pH VALUES AT VARIOUS TEMPERATURES For temperature compensation during calibration, please refer to the following chart.
EC/TDS CALIBRATION Accessories needed: • Use HI 70031 (1413 µS/cm = 1.413 mS/cm) EC calibration solution or HI 70442 (1500 ppm) TDS calibration solution. Note: The conversion between EC and TDS is made by a built-in circuit, hence it is requested to calibrate the meter only in EC or TDS range. The other range is thus automatically calibrated. PROCEDURE • Pour approximately 4 cm (1½") of a conductivity calibration solution (e.g. HI 7031) into a beaker.
PROBE MAINTENANCE PERIODIC MAINTENANCE Inspect the probe and the cable. The cable used for the connection to the meter must be intact and there must be no points of broken insulation on the cable or cracks on the probe stem or bulb. Connector must be perfectly clean and dry. If any scratches or cracks are present, replace the electrode. Rinse off any salt deposits with water.
4. if the red LED lights up again, calibration is needed (see "pH Calibration" section). Note: If it is not possible to calibrate, the probe has to be replaced with a new one. Note: For field applications, it is always recommended to keep a spare probe handy. When anomalies are not resolved with simple maintenance, change the probe and recalibrate the meter. BATTERY REPLACEMENT These meters are powered by a 9V battery that is located on the rear of the instrument.
ACCESSORIES PROBES HI 1285-0 HI 1285-5 HI 1285-6 Combination, amplified pH/EC/TDS probe with built-in temperature sensor, 8-pin DIN connector and 1 m (3.3') cable Combination, amplified pH/EC/TDS/temperature probe with built-in temperature sensor, 8-pin DIN connector and 1 m (3.3') cable Combination, amplified pH/EC/TDS/temperature probe with Cal-Check feature, built-in temperature sensor, 8pin DIN connector and 1 m (3.3') cable pH BUFFER SOLUTIONS HI 7004L HI 7006L HI 7007L HI 7009L HI 7010L pH 4.
CE DECLARATION OF CONFORMITY Recommendations for Users Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of these instruments in residential areas could cause unacceptable interferences to radio and TV equipment. The glass bulb at the end of the electrode is sensitive to electrostatic discharges. Avoid touching this glass bulb at all times.