User Guide
SPECIFICATIONS:
RANGE 0.0 to 14.0 pH
RESOLUTION 0.1 pH
ACCURACY
(@20°C/68°F) ±0.1 pH
TYPICAL EMC DEVIATION ±0.1 pH
ENVIRONMENT
0 to 50C (32 to 122F); 95% RH
CALIBRATIONCALIBRATION
CALIBRATIONCALIBRATION
CALIBRATION Manual, at two points through
offset and slope trimmers
TEMP.COMPENSATION Automatic
from 0 to 50ºC (32 to 122ºF)
BATTERY TYPE 4 x 1.5V alkaline (included)
BATTERY LIFE approx. 700 hours
of continuous use
DIMENSIONS 175 x 41 x 23 mm
(7.9 x 1.8 x 1")
WEIGHT 78 g (2.7 oz.)
ACCESSORIES:
HI 73108 Spare electrode
HI 77400P pH 4.01 & 7.01 calibration kit
(5 x 20 mL, each)
HI 77700P pH 7.01 calibration kit
(10 x 20 mL)
HI 770710P pH 7.01 & 10.01 calibration kit
(5 x 20 mL, each)
HI 7004M pH 4.01 buffer solution
(230 mL bottle)
HI 7007M pH 7.01 buffer solution
(230 mL bottle)
HI 7010M pH 10.01 buffer solution
(230 mL bottle)
HI 70300M Storage solution (230mL)
HI 7061M Electrode cleaning solution
(230mL bottle)
HI 731326 Calibration screwdriver (20 pcs)
IS98108-R1 09/01
ELECTRODE REPLACEMENT:
The electrode can be easily replaced in the following
way:
• Slide off the battery cover.
• Remove the screw on the back of the
pHep+pHep+
pHep+pHep+
pHep+
located below the battery compartment and slide
off the whole internal part of the tester.
• To remove the electrode proceed as follows:
Remove the fastening screw on the front (1) to
loose the electrode. The internal circuit is con-
nected to the electrode through four wires (one
for the glass sensor, one for the reference and
two for the temperature sensor). Remove the
two small screws (2, 3 & 4) which fasten the
four connecting wires to their sockets.
• Remove the electrode and re-
place it with a new
HI73108.
• Connect the new electrode as
follows:
first fasten the screw on the
front (1) to attach the elec-
trode to the circuit. Insert
the four small wires into their
sockets and fasten the four
screws (2, 3 & 4).
• The new electrode is supplied
with four new O-rings (5). Make sure they are
placed properly (as shown on the above draw-
ing) before reinserting the circuit into the
plastic casing.
• Reinsert the internal circuit with the new
electrode into the plastic casing.
• Fasten the screw on the back below the
battery compartment.
Internal circuit and electrode:
• Reinsert the battery compartment cover.
• Perform a new calibration before using the
pHep+ pHep+
pHep+ pHep+
pHep+ again.
pHep
®®
®®
®
+
HI 98108
Pocket-sized pH Meter
with Automatic
Temperature Compensation
HI73108
ONE YEAR WARRANTY & SERVICEABLE:ONE YEAR WARRANTY & SERVICEABLE:
ONE YEAR WARRANTY & SERVICEABLE:ONE YEAR WARRANTY & SERVICEABLE:
ONE YEAR WARRANTY & SERVICEABLE:
Covered by
one yearone year
one yearone year
one year warranty against defects in
workmanship and materials, this tester is now completely
serviceable. Contact your dealer for further information.
The
pHep+ pHep+
pHep+ pHep+
pHep+ is in compliance with the CE directives EN
50082-1, EN 50081-1 and EN 61010-1.
pHeppHep
pHeppHep
pHep
®®
®®
®
is a registered Trademark of "Hanna Instruments"