User Guide
Visit our Internet Home Page:
http://www.hannainst.com
SPECIFICATIONS:SPECIFICATIONS:
SPECIFICATIONS:SPECIFICATIONS:
SPECIFICATIONS:
RANGERANGE
RANGERANGE
RANGE 0.0 to 14.0 pH
RESOLUTIONRESOLUTION
RESOLUTIONRESOLUTION
RESOLUTION 0.1 pH
ACCURACY ACCURACY
ACCURACY ACCURACY
ACCURACY
(@20C/68F)(@20C/68F)
(@20C/68F)(@20C/68F)
(@20C/68F)
±0.2 pH
TYPICAL EMC DEVIATIONTYPICAL EMC DEVIATION
TYPICAL EMC DEVIATIONTYPICAL EMC DEVIATION
TYPICAL EMC DEVIATION ±0.1 pH
CALIBRATIONCALIBRATION
CALIBRATIONCALIBRATION
CALIBRATION Manual, at 1 point through trimmer
ENVIRONMENTENVIRONMENT
ENVIRONMENTENVIRONMENT
ENVIRONMENT 0 to 50C
(32 to 122F); 95% RH
BATTERY TYPEBATTERY TYPE
BATTERY TYPEBATTERY TYPE
BATTERY TYPE 4 x 1.5V alkaline (included)
BATTERY LIFEBATTERY LIFE
BATTERY LIFEBATTERY LIFE
BATTERY LIFE approx. 800 hours
of continuous use
DIMENSIONSDIMENSIONS
DIMENSIONSDIMENSIONS
DIMENSIONS 175 x 41 x 23 mm
(7.9 x 1.8 x 1")
WEIGHTWEIGHT
WEIGHTWEIGHT
WEIGHT 78 g (2.7 oz.)
ACCESSORIES:ACCESSORIES:
ACCESSORIES:ACCESSORIES:
ACCESSORIES:
HI 73106HI 73106
HI 73106HI 73106
HI 73106 Spare electrode
HI 77400PHI 77400P
HI 77400PHI 77400P
HI 77400P pH 4.01 & 7.01 calibration kit
(5 x 20 mL, each)
HI 77700PHI 77700P
HI 77700PHI 77700P
HI 77700P pH 7.01 calibration kit
(10 x 20 mL)
HI 770710PHI 770710P
HI 770710PHI 770710P
HI 770710P pH 7.01 & 10.01 calibration kit
(5 x 20 mL, each)
HI 7004MHI 7004M
HI 7004MHI 7004M
HI 7004M pH 4.01 buffer solution
(230 mL bottle)
HI 7007MHI 7007M
HI 7007MHI 7007M
HI 7007M pH 7.01 buffer solution
(230 mL bottle)
HI 7010MHI 7010M
HI 7010MHI 7010M
HI 7010M pH 10.01 buffer solution
(230 mL bottle)
HI 70300MHI 70300M
HI 70300MHI 70300M
HI 70300M Storage solution (230mL)
HI 7061MHI 7061M
HI 7061MHI 7061M
HI 7061M Electrode cleaning solution
(230mL bottle)
HI 731326HI 731326
HI 731326HI 731326
HI 731326 Calibration screwdriver (20 pcs)
IS98106-R3 09/01
ELECTRODE REPLACEMENT:ELECTRODE REPLACEMENT:
ELECTRODE REPLACEMENT:ELECTRODE REPLACEMENT:
ELECTRODE REPLACEMENT:
The electrode can be easily replaced in the following
way:
• Slide off the battery cover.
• Remove the screw on the back of the
ChampChamp
ChampChamp
Champ
located below the battery compartment and slide
off the whole internal part of the tester.
• To remove the electrode proceed as follows:
Remove the fastening screw on the front (1) to
loose the electrode. The internal circuit is con-
nected to the electrode through two wires (one
for the glass sensor and one for the reference).
Remove the two small screws (2 & 3) which
fasten the two connecting wires to their sock-
ets.
• Remove the electrode and
replace it with a new
HI73106.
• Connect the new electrode
as follows:
first fasten the screw on the
front (1) to attach the elec-
trode to the circuit. Insert
the two small wires into
their sockets and fasten the
two screws (2 & 3).
• The new electrode is supplied with four new O-
rings (4). Make sure they are placed properly
(as shown on the above drawing) before
reinserting the circuit into the plastic casing.
• Reinsert the internal circuit with the new
electrode into the plastic casing.
Internal circuit and electrode:
• Fasten the screw on the back below the
battery compartment.
• Reinsert the battery compartment cover.
• Perform a new calibration before using the
ChampChamp
ChampChamp
Champ
again.
ChampChamp
ChampChamp
Champ
®®
®®
®
HI 98106
Pocket-sized pH Meter
HI73106
ONE YEAR WARRANTY & SERVICEABLE:ONE YEAR WARRANTY & SERVICEABLE:
ONE YEAR WARRANTY & SERVICEABLE:ONE YEAR WARRANTY & SERVICEABLE:
ONE YEAR WARRANTY & SERVICEABLE:
Covered by
one yearone year
one yearone year
one year warranty against defects in work-
manship and materials, this tester is now completely
serviceable. Contact your dealer for further information.
The
ChampChamp
ChampChamp
Champ is in compliance with the CE directives EN
50082-1, EN 50081-1 and EN 61010-1.
Champ
®
is a registered Trademark of "Hanna Instruments"