Remove the cuvette.
For free chlorine measurement:
add 3
drops of HI 93701A-F (DPD1 indicator) and
3 drops of HI 93701B-F (DPD1 buffer) to an
empty cuvette. Swirl gently to mix, and
immediately add 10 mL of unreacted sample.
Replace the cap and shake gently.
Immediately replace the cuvette into
the cuvette holder and ensure that the
notch on the cap is positioned securely
into the groove.
Press READ/ and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
At the end of measurement, the instrument
directly displays concentration in mg/L of
free chlorine on the LCD.
For total chlorine measurement,
add 3
drops of HI 93701A-T (DPD1 indicator),
3 drops of HI 93701B-T (DPD1 buffer)
and 1 drop of HI 93701C-T (DPD3
solution) to an empty cuvette. Swirl gently
to mix, and immediately add 10 mL of
unreacted sample.
Replace the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.