Instruction Manual HI 96724C Free & Total Chlorine ISM www.hannainst.
Dear Customer, Thank you for choosing a Hanna product. This manual will provide you with the necessary information for the correct use of the instrument. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. TABLE OF CONTENTS PRELIMINARY EXAMINATION .............................................................................................. 3 GENERAL DESCRIPTION ...................................................
PRELIMINARY EXAMINATION Please examine this product carefully. Make sure that the instrument is not damaged. If any damage occurred during shipment, please notify your Dealer. Each HI 96724 Ion Selective Meter is supplied complete with: • Two Sample Cuvettes and Caps • CAL CHECK standard cuvettes • 9V Battery • Cloth for wiping cuvettes • Instrument quality certificate • Instruction Manual • Rigid carrying case Note:Save all packing material until you are sure that the instrument works correctly.
GENERAL DESCRIPTION The HI 96724 is an auto diagnostic portable microprocessor meter that benefits from Hanna’s years of experience as a manufacturer of analytical instruments. It has the advanced optical system based on a special tungsten lamp and a narrow band interference filter that allows most accurate and repeatable readings. All instruments are factory calibrated and the electronic and optical design minimizes the need of frequent calibration.
SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Light Detector Method Environment Battery Type Auto-Shut off Dimensions Weight REQUIRED REAGENTS Code HI 93701A-T HI 93701B-T HI 93701C-T 0.00 to 5.00 mg/L of Cl2 (Free or Total) 0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 mg/L above 3.50 mg/L ±0.03 mg/L ±3% of reading @ 25°C ±0.01 mg/L Tungsten lamp Silicon Photocell with narrow band interference filter @ 525 nm Adaptation of the USEPA method 330.5 and Standard Method 4500-Cl G.
PRINCIPLE OF OPERATION Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation and matter. When a light beam crosses a substance, some of the radiation may be absorbed by atoms, molecules or crystal lattices.
A microprocessor controlled special tungsten lamp emits radiation which is first optically conditioned and beamed to the sample contained in the cuvette. The optical path is fixed by the diameter of the cuvette. Then the light is spectrally filtered to a narrow spectral bandwidth, to obtain a light beam of intensity Io or I. The photoelectric cell collects the radiation I that is not absorbed by the sample and converts it into an electric current, producing a potential in the mV range.
KEYPAD DESCRIPTION • ON/OFF: to turn the meter on and off. • ZERO/CFM: to zero the meter prior to measurement, to confirm edited values or to confirm factory calibration restore. • READ/ : this is a bi-functional key. In measurement mode, press to make a measurement. In GLP mode press to view the next screen. • CAL CHECK: this is a bi-functional key. Just press to perform the validation of the meter, or press and hold for three seconds to enter calibration mode. • GLP/ : this is a bi-functional key.
ERRORS AND WARNINGS The instrument shows clear messages when erroneous condition appears. Messages are also displayed when the obtained values are outside expected range. The beeper is playing a beep on errors. a) on zero reading Light High: There is too much light to perform a measurement. Please check the preparation of the zero cuvette. Light Low: There is not enough light to perform a measurement. Please check the preparation of the zero cuvette. No Light:The instrument cannot adjust the light level.
Under range: A blinking “0.00” indicates that the sample absorbs less light than the zero reference. Check the procedure and make sure you use the same cuvette for reference (zero) and measurement. Over Range: A flashing value of the maximum concentration indicates an over range condition. The concentration of the sample is beyond the programmed range: dilute the sample and re-run the test. c) during calibration procedure Standard Low: The standard reading is less than expected.
Dead battery: This indicates that the battery is dead and must be replaced. Once this indication is displayed, the meter will lock up. Change the battery and restart the meter. GENERAL TIPS FOR AN ACCURATE MEASUREMENT The instructions listed below should be carefully followed during testing to ensure best accuracy. • Color or suspended matter in large amounts may cause interference, therefore these should be removed by treatment with active carbon and by prior filtration.
• In order to avoid reagent leaking and to obtain more accurate measurements, it is recommended to close the cuvette first with the supplied HPDE plastic stopper and then with the black cap. • Each time the cuvette is used, the cap must be tightened to the same degree. • Whenever the cuvette is placed into the measurement cell, it must be dry outside, and completely free of fingerprints, oil or dirt. Wipe it thoroughly with HI 731318 or a lint-free cloth prior to insertion.
MEASUREMENT PROCEDURE To compensate the meter for the sample turbidity or color, the measurement takes place in two phases. First, the meter is zeroed using the unreacted sample. After the reagents are added the reacted sample is measured. • Turn the meter on by pressing ON/OFF. The display briefly shows all tags on. • When the beeper sounds briefly and the LCD displays dashes, the meter is ready. The blinking “ZERO” indicates that the instrument needs to be zeroed first.
• Remove the cuvette. • For free chlorine measurement: add 3 drops of HI 93701A-F (DPD1 indicator) and 3 drops of HI 93701B-F (DPD1 buffer) to an empty cuvette. Swirl gently to mix, and immediately add 10 mL of unreacted sample. Replace the cap and shake gently. • Immediately replace the cuvette into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove.
• Wait for 2’ and 30’’ and press READ/ . The lamp, cuvette and detector icons will appear on the display, depending on the measurement phase. • At the end of measurement, the instrument directly displays concentration in mg/L of total chlorine on the LCD.
• Place the CAL CHECK TM Standard HI 96724-11 Cuvette A into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove. • Press ZERO/CFM and the lamp, cuvette and detector icons will appear on the display, depending on the measurement phase. • After a few seconds, the display will show “-0.0-”. The meter is now zeroed and ready for validation. • Remove the cuvette.
CALIBRATION PROCEDURE Note: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK or ON/OFF keys. Warning: Do not calibrate the instrument with standard solutions other than the HANNA CAL CHECKTM Standards, otherwise erroneous results will be obtained. For accurate calibration, please perform test at room temperature: 18 to 25°C (64.5 to 77.0°F) • Turn the meter on by pressing ON/OFF. • When the beeper sounds briefly and the LCD displays dashes, the meter is ready.
• After a few seconds the display will show “-0.0-”. The meter is now zeroed and ready for calibration. The blinking “READ” asks for reading calibration standard. • Remove the cuvette. • Place the CAL CHECKTM Standard HI 96724-11 Cuvette B into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove. • Press READ/ and the lamp, cuvette and detector icons will appear on the display, depending on the measurement phase.
DATE INPUT • Press GLP/ to edit the desired year (2000-2099). If the key is kept pressed, the year number is automatically increased. or • When the correct year has been set, press ZERO/CFM or READ/ to confirm. Now the display will show the month blinking. • Press GLP/ to edit the desired month (01-12). If the key is kept pressed the month number is automatically increased. or • When the correct month has been set, press ZERO/CFM or READ/ to confirm. Now the display will show the day blinking.
• The instrument will return automatically to the measurement mode by displaying dashes on the LCD. GLP In the GLP mode, the last user calibration date can be consulted and the factory calibration can be restored. LAST CALIBRATION DATE To display the calibration date: • Press GLP/ to enter GLP mode. The calibration month and day will appear on the main display and the year on the secondary display. • If no calibration was performed, the factory calibration message, “F.
• Press READ/ to enter in the factory calibration restore screen. The instrument asks for confirmation of user calibration delete. • Press ZERO/CFM to restore the factory calibration or press GLP/ again to abort factory calibration restore. • The instrument briefly notifies “done” when restores factory calibration and returns to measurement mode. BATTERY MANAGEMENT To save battery, the instrument shuts down after 10 minutes of non-use in measurement mode and after 1 hour of non-use in calibration mode.
BATTERY REPLACEMENT To replace the instrument’s battery, follow the steps: • Turn the instrument off by pressing ON/OFF. • Turn the instrument upside down and remove the battery cover by turning it counterclockwise. • Extract the battery from its location and replace it with a fresh one. • Insert back the battery cover and turn it clockwise to close.
WARRANTY HI 96724 is warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained according to the instructions. This warranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact your dealer. If under warranty, report the model number, date of purchase, serial number and the nature of the failure.