Owner's manual
5
SPECIFICATIONS
PRECISION AND ACCURACY
Range 0.00 to 5.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.04 mg/L ± 2% of reading @ 25°C
Typical EMC Deviation ±0.01 mg/L
Light Source Tungsten lamp
Light Detector Silicon Photocell with narrow band interference filter @ 525 nm
Method Adaptation of the USEPA method 315B and Standard Method 3500-Fe B.
The reaction between iron and the phenantroline reagent causes an
orange tint in the sample.
Environment 0 to 50°C (32 to 122°F); max 95% RH non-condensing
Battery Type 1 x 9 volt
Auto-Shut off After 10' of non-use in
measurement mode
;
after 1 hour of non-use in
calibration mode
;
with last reading reminder.
Dimensions 192 x 102 x 67 mm (7.6 x 4 x 2.6")
Weight 290 g (10 oz.).
REQUIRED REAGENTS
Code Description Quantity/test
HI 93721-0 Iron High Range Powder Reagent 1 packet
PrecisionPrecision
PrecisionPrecision
Precision is how closely repeated measurements
agree with each other. Precision is usually
expressed as standard deviation (SD).
AccuracyAccuracy
AccuracyAccuracy
Accuracy is defined as the nearness of a test
result to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure
explains these definitions.
In a laboratory using a standard solution of 1.50
mg/L Iron and a representative lot of reagent, an
operator obtained with a single instrument a
standard deviation of 0.010 mg/L.