15
Calcium (Ca
2+
): above 10,000 mg/L as CaCO
3
Copper (Cu
2+
): does not interfere: reagent contains a masking
agent.
Magnesium (Mg
2+
): above 100,000 mg/L as CaCO
3
Molybdate Molybdenum: above 25 mg/L (as Mo-MoO
4
) (positive error)
Sample containing up to 50 mg/L Mo-MoO
4
may
be measured if a reading is taken after 1 minute
instead of 3 minutes.
High Iron concentrations: up to 200 mg/L do not inhibit color development;
above 200 mg/L color development may be inhibited.
High sulfide concentrations (S
2-
): high sulfide concentrations may interfere
To reduce sulfide: acidify the sample and boil for
about 20 minutes, neutralize and adjust volume
before measurements.
Silica (SiO
2
): above 175 mg/L (as SiO
2
). May slow down color
development (negative error)
VALIDATION PROCEDURE
Use the validation procedure to ensure that the instrument is properly calibrated.
Warning: Do not validate the instrument with any standard solutions other than the HANNA CAL
CHECK
TM
Standards, otherwise erroneous results will be obtained.
Turn the meter on by pressing ON/OFF.
Place the CAL CHECK
TM
Standard
HI 96721-11 Cuvette A into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.