Owner's manual

14
Remove the cuvette.
Add the content of one packet of
HI 93721-0 Iron High Range reagent to
the cuvette. Replace the cap and shake
gently until dissolution is complete.
Replace the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
Press and hold READ/ /TIMER for three
seconds. The display will show the
countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.
Alternatively, wait for 3 minutes and just
press READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.
At the end of measurement, the instrument
directly displays concentration in mg/L of
iron on the LCD.
INTERFERENCES
Alkalinity: above 2,000 mg/L (as CaCO
3
). May slow down
color development (negative error).
To resolve this, neutralize the sample with diluted
HCl