Dear Customer, Thank you for choosing a Hanna product. This manual will provide you with the necessary information for the correct use of the instrument. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. TABLE OF CONTENTS PRELIMINARY EXAMINATION .............................................................................................. 3 GENERAL DESCRIPTION ...................................................
GENERAL DESCRIPTION The HI 96720 is an auto diagnostic portable microprocessor meter that benefits from Hanna’s years of experience as a manufacturer of analytical instruments. It has the advanced optical system based on a special tungsten lamp and a narrow band interference filter that allows most accurate and repeatable readings. All instruments are factory calibrated and the electronic and optical design minimizes the need of frequent calibration.
PRINCIPLE OF OPERATION Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation and matter. When a light beam crosses a substance, some of the radiation may be absorbed by atoms, molecules or crystal lattices.
KEYPAD DESCRIPTION • ON/OFF: to turn the meter on and off. • ZERO/CFM: to zero the meter prior to measurement, to confirm edited values or to confirm factory calibration restore. • READ/ : this is a bi-functional key. In measurement mode, press to make a measurement. In GLP mode press to view the next screen. • CAL CHECK: this is a bi-functional key. Just press to perform the validation of the meter, or press and hold for three seconds to enter calibration mode. • GLP/ : this is a bi-functional key.
Under range: A blinking “0.00” indicates that the sample absorbs less light than the zero reference. Check the procedure and make sure you use the same cuvette for reference (zero) and measurement. Battery low: The battery must be replaced soon. Dead battery: This indicates that the battery is dead and must be replaced. Once this indication is displayed, the meter will lock up. Change the battery and restart the meter.
• It is important that the sample does not contain any debris. This would corrupt the reading. • Each time the cuvette is used, the cap must be tightened to the same degree. • Whenever the cuvette is placed into the measurement cell, it must be dry outside, and completely free of fingerprints, oil or dirt. Wipe it thoroughly with HI 731318 or a lint-free cloth prior to insertion. • Shaking the cuvette can generate bubbles in the sample, causing higher readings.
Now, follow normal measurement procedure. The true value of the sample is the reading obtained multiplied by a factor of one hundred (the dilution factor). For your reference, factors to convert readings in mg/L to French degrees (FD), German degrees (DD) and English degrees (ED) of hardness are as follows: • After a few seconds, the display will show “-0.0-”. The meter is now zeroed and ready for measurement.
• After a few seconds, the display will show “-0.0-”. The meter is now zeroed and ready for validation. • Remove the cuvette. CALIBRATION PROCEDURE Note: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK or ON/OFF keys. Warning: Do not calibrate the instrument with standard solutions other than the HANNA CAL CHECKTM Standards, otherwise erroneous results will be obtained. • Turn the meter on by pressing ON/OFF.
• After a few seconds the display will show “-0.0-”. The meter is now zeroed and ready for calibration. The blinking “READ” asks for reading calibration standard. DATE INPUT • Press GLP/ to edit the desired year (2000-2099). If the key is kept pressed, the year number is automatically increased. • Remove the cuvette. • Place the CAL CHECKTM Standard HI 96720-11 Cuvette B into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove.
• The instrument will return automatically to measurement mode by displaying dashes on the LCD. • Press READ/ to enter in the factory calibration restore screen. The instrument asks for confirmation of user calibration delete. GLP In the GLP mode, the last user calibration date can be consulted and the factory calibration can be restored. LAST CALIBRATION DATE To display the calibration date: • Press GLP/ to enter GLP mode.
BATTERY REPLACEMENT To replace the instrument’s battery, follow the steps: • Turn the instrument off by pressing ON/OFF. • Turn the instrument upside down and remove the battery cover by turning it counterclockwise. • Extract the battery from its location and replace it with a fresh one. WARRANTY HI 96720 is warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained according to the instructions.