Manual

14 15
Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
• After a few seconds, the display will show
“-0.0-”. The meter is now zeroed and
ready for measurement.
Remove the cuvette.
Add 10 drops of HI 93717A-0 Phosphate
reagent A to the cuvette.
Add the content of one packet of
HI 93717B-0 Phosphate reagent B to
the cuvette. Replace the cap and shake
gently until dissolution is complete.
Replace the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
Press and hold
READ/READ/
READ/READ/
READ/
/TIMER/TIMER
/TIMER/TIMER
/TIMER for three
seconds. The display will show the
countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.
Alternatively, wait for 5 minutes and just
press
READ/READ/
READ/READ/
READ/
/TIMER/TIMER
/TIMER/TIMER
/TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.
At the end of measurement, the instrument
directly displays concentration in mg/L of
phosphate on the LCD.
INTERFERENCES
Calcium (Ca
2+
): above 10,000 mg/L as CaCO
3
Chloride (Cl
-
): above 150,000 mg/L as Cl
-
Ferrous iron (Fe
2+
): above 100 mg/L as Fe
2+
Magnesium (Mg
2+
): above 40,000 mg/L as CaCO
3
Sulfide (S
2+
)
VALIDATION PROCEDURE
Use the validation procedure to ensure that the instrument is properly calibrated.
Warning: Do not validate the instrument with any standard solutions other than the HANNA CAL
CHECK
TM
Standards, otherwise erroneous results will be obtained.
Turn the meter on by pressing ON/OFF.
Place the CAL CHECK
TM
Standard
HI 96717-11 Cuvette A into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.