User guide

14
Remove the cuvette.
Add the content of one packet of
HI 93707-0 Nitrite Low Range reagent.
Replace the cap and shake gently for
15 seconds.
Replace the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
Press and hold READ/ /TIMER for three
seconds. The display will show the
countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.
Alternatively, wait for 15 minutes and
just press READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.
At the end of measurement, the
instrument directly displays concentration
in mg/L of nitrogen-nitrite (NO
2
¯-N) on
the LCD.
To convert the NO
2
¯-N concentration to
the nitrite ion concentration (NO
2
¯),
multiply the reading by factor of 3.29.
To convert the NO
2
¯-N concentration to
sodium nitrite concentration (NaNO
2
),
multiply the reading by factor of 4.93.