User Guide
36
TM? Causes the meter to send the tem-
perature value. If the reading is out of
range "Err 3" is sent.
DA? Requests the meter to send the date
E.g. "022896" for 28th Feb. 96
TI? Requests the meter to send the time
E.g. "233001"
for 23:30 hr, 1 sec. as interval
"233002"
for 23:30 hr, 15 sec. as interval
"233003"
for 23:30 hr, 30 sec. as interval
"233004"
for 23:30 hr, 1 min. as interval
"233005"
for 23:30 hr, 5 min. as interval
"233006"
for 23:30 hr, 30 min. as interval
"233007"
for 23:30 hr, 60 min. as interval
"233008"
for 23:30 hr, 120 min. as interval
"233009"
for 23:30 hr, 180 min. as interval
?ML Requests the meter to send the avail-
able lot number collected in memory.
The transmission begins with <STX>
and terminates with <ETX>. The data
are sent in the following order:
1) stx
2) Lot number
E.g. "01" for lot No. 1
3) Total number of samples per lot
E.g. "1234" for total no. of
samples: 1234.