User Guide

18
minutes to ensure precise
conditioning of the probe.
Remove the protective cap
from the D.O. probe.
Set the appropriate altitude
factor (see page 23). Make
sure the salinity factor is set
to zero (see page 25).
ZERO CALIBRATION
Dip the probe into HI 7040
zero oxygen solution and
stir gently for 2-3 minutes.
Press the CAL key and the "BUF" indica-
tor will blink until the reading is stable.
As soon as the reading is stable, the
"CFM" indicator will start blinking. Press
the CFM key to confirm the "0.0%" D.O.
reading.
If the reading is not close to the selected
value, "WRONG
BUF
1
" and "WRONG " will
blink alternatively.
CAL
%
BUF
%
CFM
%
WRONG
BUF
%
WRONG
CFM