User Guide

9
FUNCTIONAL DESCRIPTION &FUNCTIONAL DESCRIPTION &
FUNCTIONAL DESCRIPTION &FUNCTIONAL DESCRIPTION &
FUNCTIONAL DESCRIPTION &
SPECIFICATIONS OF HI 933100SPECIFICATIONS OF HI 933100
SPECIFICATIONS OF HI 933100SPECIFICATIONS OF HI 933100
SPECIFICATIONS OF HI 933100
1) Battery Compartment
2) Probe Socket
3) Liquid Crystal Display
4) ON/OFF Key
5) CAL, to enter or exit cali-
bration mode
6) FACTOR, to set the tem-
perature coefficient or select
the temperature mode
7) 12VDC Power Socket
8) 'LOBAT' Low Battery
9) CFM, to confirm the cali-
bration reading
10) BUF, to select the calibra-
tion buffer value
11) DOWN, to set the tempera-
ture coefficient or for manual temperature compensation setting
12) UP, to set the temperature coefficient or for manual tempera-
ture compensation setting
2
7
1
4
6
8
10
11
12
9
5
3
ins
rum n
s
CAL
ON/OFF
FACTOR UP DOWN
CFM BUF
HI 933100
MICROPROCESSOR
CONDUCTIVITY METER
CONDUCTIVITY
mS
LO BAT
°C
Range Conductivity 0.0 to 150.0 µS/cm
150 to 1500 µS/cm
1.50 to 15.00 mS/cm
15.0 to 199.9 mS/cm
Temperature 0.0 to 60.0°C
Resolution 0.1 µS/cm; 1 µS/cm; 0.01 mS/cm;
0.1 mS/cm; 0.1°C
Accuracy ±1% Full Scale; ±0.5°C
(@20°C/68°F) excluding probe error
Typical EMC ±1% Full Scale
Deviation ±0.2 °C
Calibration Automatic, two points
at 0 and 1413 or 12880 µS/cm
Temperature Automatic or manual from 10 to 60°C
Compensation (50 to 140°F) with ß from 0 to 3%/°C
Probe HI 76302W with 1 m (3.3') cable
(included)
Battery Type/Life 4 x 1.5V AA; 150 hours of
continuous use. Power socket for 12VDC
Environment 0 to 50°C (32 to 122°F); 100% RH
Dimensions 196 x 80 x 60 mm (7.7 x 3.1 x 2.2")
Weight 425 g (15 oz.); Kit: 1.4 kg (3.1 lb.)