User Guide

28
Immerse the conductivity probe into the
50°C (122°F) bath and wait for a couple
of minutes for thermal equilibrium.
Press CAL to display the "50.0°C", "CAL" symbol and an intermit-
tent "BUF" indicator. " " will symbol blink for a few seconds.
" " will disappear from the display
when the reading is stable and "CON" will
start blinking.
Press CFM to confirm the calibration value.
The calibration is now complete and the instrument returns to normal
operation.
Note: You can quit the calibration mode at any
time by pressing ON/OFF.
°C
50 °C
(122 °F)
CAL
BUF
CAL
°C
°C
CON
CFM
ON/OFF