User Guide
28
• Immerse the conductivity probe into the
50°C (122°F) bath and wait for a couple
of minutes for thermal equilibrium.
• Press CAL to display the "50.0°C", "CAL" symbol and an intermit-
tent "BUF" indicator. " " will symbol blink for a few seconds.
• " " will disappear from the display
when the reading is stable and "CON" will
start blinking.
• Press CFM to confirm the calibration value.
The calibration is now complete and the instrument returns to normal
operation.
Note: You can quit the calibration mode at any
time by pressing ON/OFF.
°C
50 °C
(122 °F)
CAL
BUF
CAL
°C
°C
CON
CFM
ON/OFF