User Guide

14
Select the temperature (°C) mode by pressing
FACTOR once.
The secondary LCD figure will display a default
temperature of 25°C or the last recorded tem-
perature reading with a blinking "°C" symbol.
Change the reading by pressing the UP or
DOWN keys. Hold down the UP or DOWN keys
to force the reading to change in 0.5°C steps
(HI 933100) or 1°C (HI 9635) instead of
0.1°C.
Note: If HI 76302W is connected to HI 933100
or HI 9635 and the °C mode is se-
lected, the secondary LCD will continue
to display the measured temperature ,
and the "°C" symbol will not blink.
Pressing UP or DOWN will also not alter
the temperature readings.
TEMPERATURE COEFFICIENT SELECTION (FOR
HI 933100 AND HI 9635)
Press FACTOR again to select the TC (Temperature Coefficient). The
secondary LCD will display the temperature coefficient (e.g. 2.0% per
°C).
This value is adjustable from 0.0 to 3.0%/°C
by pressing UP or DOWN.
AFTER MEASUREMENTS
When all measurements are completed the meter
should be turned OFF and the probe cleaned (see
page 30).
Note: The probe body and sleeve are made of PVC and are
susceptible to damage due to temperatures exceeding 60°C
(140°F). If the probe is exposed to high temperature, the
bond between the rings and the probe body may become
impaired and the probe will not function properly, in which
case, it has to be replaced.