User Guide

25
Rinse the electrode with
some pH 10.01 (or pH 9.18)
solution. Dip the bottom 4
cm (1½") of the electrode
in a beaker containing the
third buffer.
When the CFM indicator starts blinking,
indicating that the electrode has stabilized,
press the ALT and CFM keys to confirm
the third calibration point.
If the buffer is out of the slope range,
"WRONG " will blink. Press the ALT and
CAL keys simultaneously and recalibrate
the value.
If you cannot complete the calibration,
check your electrode by following the con-
ditioning and maintenance procedure on
page 30 and repeat the calibration proce-
dure from the beginning with fresh buffers.
The pH electrode might have to be replaced
if calibration cannot still be successfully
performed.
If everything is satisfactory the calibration
ends and the meter will return to the pH
measurement mode.
4 cm
(1½
")
HI 7010
CFM
ALT
CFM
ALT
CAL
WRONG
BUF