HANNA LITERATURE Instruction Manual Hanna publishes a wide range of catalogs and handbooks for an equally wide range of applications. The reference literature currently covers areas such as: HI 8014 - HI 8314 HI 8424 - HI 8915 HI 9214 - HI 931000 HI 9622 Portable pH Meters • Water Treatment • Process • Swimming Pools • Agriculture • Food • Laboratory • Thermometry and many others. New reference material is constantly being added to the library.
Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the meter. This manual will provide you with the necessary information for a correct use of the instrument, as well as a more precise idea of its versatility. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. These instruments are in compliance with the directives. CE DECLARATION OF CONFORMITY TABLE OF CONTENTS PRELIMINARY EXAMINATION ...........
OTHER PRODUCTS FROM HANNA • • • • • • • • • • • • • • • • • • • CALIBRATION AND MAINTENANCE SOLUTIONS CHEMICAL TEST KITS CHLORINE METERS CONDUCTIVITY/TDS METERS DISSOLVED OXYGEN METERS HYGROMETERS ION SPECIFIC METERS (Colorimeters) MAGNETIC STIRRERS Na/NaCl METERS pH/ORP/Na ELECTRODES PROBES (DO, µS/cm, RH, T, TDS) PUMPS REAGENTS SOFTWARE THERMOMETERS TITRATORS TRANSMITTERS TURBIDITY METERS Wide Range of Accessories Most Hanna meters are available in the following formats: • BENCH-TOP METERS • POCKET-SIZE
• • • • measurements with automatic range switching when readings exceed ±399.9 mV. HI 8424 comes supplied with: HI 1230B combination, double-junction, gel pH electrode, HI 7669AW temperature probe and a 9V battery. HI 8915 is a portable pH/mV/°C meter that combines accurate pH measurements with a recorder output capability and automatic temperature compensation. Two recorder output terminals make it possible to attach a plotter or recorder directly to the meter for a hard copy of the measurements.
ELECTRODE APPLICATION REFERENCE GUIDE Application 1. Aquarium 2. Bath-water 3. Beer 4. Bread 5. Cheese 6. Dairy products 7. Dirty water 8. Emulsions 9. Environment 10. Flasks 11. Food industry general use 12. Fruit 13. Fruit juices, organic 14. Galvanizing waste solution 15. High purity water 16. Horticulture 17. Laboratory general use 18. 19. 20. 21. 22. 23. 24. 25. 26. 27. 28. 29. 30. 31. 32. 33. 34. 35. 36. 37. 38. 39. 40.
SPECIFICATIONS HI 8014 HI 8014 Range pH 0.00 to 14.00 mV ±1999 Resolution pH 0.01 mV 1 Accuracy pH ±0.01 (@20°C/68°F) mV ±1 Typical EMC pH ±0.05 Deviation mV ±8 pH Calibration Manual 2 point through offset and slope trimmers Offset Calibration ±1 pH Slope Calibration from 85 to 105% Temperature Manual from 0 to 100°C Compensation (32 to 212°F) Electrode HI 1332B combination, double-junction, refillable pH electrode with BNC connector and 1 m (3.
HI 1413B Glass-body, single junction, flat tip, Viscolene, non-refillable combination pH electrode. Use: surface measurement. 12 mm 0.5" HI 1413 FUNCTIONAL DESCRIPTION HI 8314 5 1 2 pH 110 mm 4.3" ORP ELECTRODES HI 3131B / HI 3111S Glass-body, refillable, combination platinum ORP electrode. Use: titration. HI 8314 membrane pHmeter ON OFF 6 pH 7 °C 8 12 mm 0.5" HI 3131 3 150 mm 5.9" "S" VERSION HI 3230B / HI 3210S Plastic-body (Ultem®), gel-filled, combination platinum ORP electrode.
SPECIFICATIONS HI 8314 Range pH mV °C Resolution pH mV °C Accuracy pH (@20°C/68°F) mV °C Typical EMC pH Deviation mV °C pH Calibration Offset Calibration Slope Calibration Temperature Compensation Electrode Temperature Probe Input Impedance Battery Type Life Environment Dimensions Weight HI 8314 0.00 to 14.00 ±1999 0.0 to 100.0 0.01 1 0.1 ±0.01 ±1 ±0.4 ±0.
HI 1330B / HI 1310S Glass-body, semimicro, single junction, refillable, combination pH electrode. Use: laboratory. 5mm DIA 0.2" FUNCTIONAL DESCRIPTION HI 8915 7 1 5mm 0.2" 2 HI 1330 pH 120 mm 4.7" "S" VERSION HI 1331B / HI 1311S Glass-body, semimicro, single junction, refillable, combination pH electrode. Use: flasks. 8 mm 0.3" 7.5mm DIA 0.29" HI 1331 HI 8915 ATC pHmeter ON OFF 8 3 HOLD pH 9 4 mV °C 10 OFFSET 5 6 210 mm 8.
SPECIFICATIONS HI 8915 pH ELECTRODES B = BNC CONNECTION PLUG + 1 m (3.3’) CABLE HI 8915 Range pH 0.00 to 14.00 mV ±1999 °C 0.0 to 100.0 Resolution pH 0.01 mV 1 °C 0.1 Accuracy pH ±0.01 (@20°C/68°F) mV ±1 °C ±0.5 Typical EMC pH ±0.02 Deviation mV ±1 °C ±0.
FUNCTIONAL DESCRIPTION HI 9214 CFM 1 REFILLING ELECTROLYTE SOLUTIONS HI 7071 HI 7072 HI 7082 CAL ON OFF General Cleaning Solution, 230 mL General Cleaning Solution, 460 mL Protein Cleaning Solution, 230 mL Protein Cleaning Solution, 230 mL Oil & Fat Cleaning Solution, 230 mL Oil & Fat Cleaning Solution, 460 mL HI 9214 stick pHmeter HI 8061M HI 8061L HI 8073M HI 8073L HI 8077M HI 8077L 2 ins rum n s ELECTRODE CLEANING SOLUTIONS IN FDA APPROVED BOTTLE 3 4 5 pH Inorganic Cleaning Solution, 230 mL
SPECIFICATIONS HI 9214 ACCESSORIES pH CALIBRATION SOLUTIONS Range Resolution Accuracy (@20°C/68°F) Typical EMC Deviation pH Calibration Offset Calibration Slope Calibration Temperature Compensation Electrode Input Impedance Battery Type Life Environment Dimensions Weight pH °C pH °C pH °C pH °C HI 9214 0.00 to 14.00 0.0 to 70.0 0.01 0.1 ±0.02 ±0.5 ±0.08 ±1 Automatic 2 point with 3 memorized standard buffers (pH 7.01, 4.01 or 10.
FUNCTIONAL DESCRIPTION HI 8424 DISPLAY CODE GUIDE HI 8014, HI 8314, HI 8424, HI 8915, HI 9214 AND HI 9622 °C Symbol indicates the meter is in temperature mode. A flashing °C symbol in HI 8424 and HI 9622 means that the temperature probe is not connected or the cable is damaged pH Symbol indicates the meter is in pH mode. HI 8014, HI 8314, HI 8424, HI 8915 AND HI 9622 mV Symbol indicates the meter is in mV mode.
SPECIFICATIONS HI 8424 Range pH mV °C pH mV °C Accuracy pH (@20°C/68°F) mV °C Typical EMC pH Deviation mV °C pH Calibration Resolution Offset Calibration Slope Calibration Temperature Compensation Electrode Temperature Probe Input Impedance Battery Type Life Environment Dimensions Weight HI 8424 0.00 to 14.00 ±399.9 mV ±400 to ±1999 mV 0.0 to 100.0 0.01 0.1mV / 1mV 0.1 ±0.01 ±0.2mV / ±1mV ±0.4 ±0.02 ±0.2 / ±1 ±0.4 Automatic 2 point with 3 memorized standard buffers (pH 7.01, 4.01 or 10.01) ±1.
Note: For field applications, it is always recommended to keep a spare electrode handy. When anomalies are not resolved with a simple maintenance, change the electrode (and recalibrate the meter) to see if the problem is alleviated. FUNCTIONAL DESCRIPTION HI 931000 1 ins rum n s 2 TAKING REDOX MEASUREMENTS HI 8014, HI 8314, HI 8424, HI 8915 and HI 9622 have the capability to take ORP measurements. An optional ORP electrode must be used to perform these measurements.
SPECIFICATIONS HI 931000 Range Resolution Accuracy (@20°C/68°F) Typical EMC Deviation pH Calibration Temperature Compensation Electrode Battery Type Life Environment Dimensions Weight pH °C pH °C pH °C pH °C HI 931000 0.00 to 14.00 0.0 to 70.0 0.01 0.1 ±0.01 ±0.5 ±0.30 ±1 Automatic 2 point with 3 memorized standard buffers (pH 7.01, 4.01, 10.
If the bulb and/or junction are dry, soak the electrode in HI 70300 Storage Solution for at least one hour. For refillable electrodes: If the refill solution (electrolyte) is more than 2½ cm (1") below the fill hole, add HI 7082 3.5M KCl Electrolyte Solution for double junction or HI 7071 3.5M KCl+AgCl Electrolyte Solution for single junction electrodes. For AmpHel® electrodes: If the electrode does not respond to pH changes, the battery is run down and the electrode should be replaced.
SPECIFICATIONS HI 9622 Range pH ISE ORP °C Resolution pH ISE ORP °C Accuracy pH (@20°C/68°F) ISE ORP °C Typical EMC pH Deviation ISE ORP °C pH Calibration Offset Calibration Slope Calibration Temperature Compensation Electrode Temperature Probe Input Impedance Battery Type Life Environment Dimensions Weight ELECTRODE CONDITIONING AND MAINTENANCE HI 9622 0.00 to 14.00 ±399.9 mV ±400 to ±1999 mV 0.0 to 100.0 0.01 0.1mV 1mV 0.1 ±0.01 ±0.2mV ±1mV ±0.4 ±0.01 ±0.2mV ±1mV ±0.
SYMPTOMS For HI 8424 only Display shows "E5" PROBLEM SOLUTION Erroneous buffer so- Use fresh buffer solution lution used for slope calibration An out of order pH Replace the electrode electrode For HI 9622 only Display shows "pH----" Out of range- pH a) Check the calibration scale b) Ensure that the pH value is between 0 and 14 c) Check the level of electrolyte and the state of the electrode itself Display shows "°C" No temperature Make sure the temperature probe is attached properly probe attached bli
For HI 931000: Attach the pH/°C electrode to the DIN socket on the top of the meter, tighten the retainer ring and turn the meter on by pressing ON/OFF. ON ins rum n s pH OFF ON OFF °C CAL CFM HI 931000 TAKING pH MEASUREMENTS °C 4 cm (1½") For HI 8014: To take a pH measurement simply submerge the tip (4cm/1½") of the electrode and a ChecktempC (or an accurate thermometer) in the sample to be tested. Turn the rotary knob to the °C position to display temperature setting on the LCD.
TEMPERATURE CALIBRATION mV OUTPUT It is possible to attach the HI 8915 to a recorder or plotter using the mV output sockets. pH, mV and °C readings can be transmitted through the output sockets at a voltage that proportionally matches the displayed value. For example, when the meter reads 2.00, 7.00 or 12.00 pH, it emits a 200, 700 or 1200 mV signal, respectively. This makes it unnecessary to use conversion tables and charts.
compensate for it (HI 8424 and HI 9622 only). In this case, the temperature probe must be disconnected. HI 8014 provides only for manual compensation. With HI 8314 and HI 8915 when the temperature probe is disconnected, the pH reading will be compensated at 25°C. If a different compensation is required, reattach the probe. • Use the UP and DOWN keys to set the temperature manually to the above value.
pH CAL To measure the mV of a sample submerge the ORP electrode tip (4 cm/1½") in the sample to be tested. Allow a few minutes for the readings to stabilize. Also, see page 42 for more information about REDOX measurements. Note (for HI 8424 and HI 9622): the resolution is 0.1 mV for values within the ±400mV range. 4 cm (1½") • The meter will request pH 4.01 or 10.01 buffer by showing "Er-4".
HOLD FUNCTION (HI 8915 ONLY) • Press RANGE to display pH measurement. The hold function is activated by the HOLD key. The measured value is frozen on the display when this key is pressed. "S" symbol will appear on the display to indicate the holding (storage) mode. Press the HOLD key again to return to regular operational mode. • Press CAL to start calibration and to display the last calibration date again.
• If the ON/OFF key is pressed at this point, the calibration process will end with only the offset (one-point) calibration. For best accuracy, however, it is recommended that a two-point calibration is performed. • The LCD will now blink "4.01 ∆". pH 10.01 can be selected by pressing the CAL key. ∆ • Rinse with pH 4.01 (or pH 10.01) buffer solution (2nd calibration point). • Dip the electrode into a pH 4.01 (or pH 10.01) buffer solution.
CFM eter k pHm 4 stic • Rinse the pH electrode with some pH 4.01 or 10.01 buffer solution (2nd calibration point). • Dip the electrode into the pH 4.01 (or pH 10.01) buffer solution. HI 921 mometer), e.g. 20°C. • Set the rotary knob to °C to display the manual temperature setting. • Adjust the °C trimmer until the LCD shows the measured temperature. HI 7004 • Stir gently and wait until the stability indicator "∆" stops flashing.
• Remove the protective cap from the electrode, rinse the tip of the electrode with some pH 7.01 solution, then immerse the pH electrode and temperature probe in a pH 7.01 buffer solution. Stir gently and wait a couple of minutes for thermal equilibrium. Note: the electrode should be submerged approximately 4 cm (1½") into the solution. The temperature probe should be located close to the pH electrode. • Press °C to display the temperature of the buffer (e.g. 20.0°C).
PROCEDURE FOR HI 8424 ON • Press RANGE to display pH measurement. • Rinse the electrode and temperature probe before immersing them into pH 4.01 or pH 10.01 calibration buffer. Stir gently and wait for thermal equilibrium. • The symbol "E5" should disappear and a temperature compensated buffer value starts blinking. • Wait until the pH symbol stops flashing. HI 7007 4 cm (1½") • Switch the meter on after connecting the pH electrode and the temperature probe.