User guide
5
Range 0.000 to 1.500 mg/L
Resolution 0.001 mg/L for
measurement
0.01 mg/L for
calibration
and
validation
Precision ±0.015 mg/L @ 0.750 mg/L
Typical EMC Deviation ±0.001 mg/L
Light Source Tungsten lamp with narrow band interference filter @ 560 nm
Light Detector Silicon Photocell
Method Adaptation of the USEPA approved bicinchoninate method. The reaction
between Copper and the bicinchoninate reagent causes a purple tint in
the sample.
Environment 0 to 50°C (32 to 122°F); max 95% RH non-condensing
Battery Type 1 x 9 volt
Auto-Shut off
After 10' of non-use in
measurement mode
;
after 1 hour of non-use in
calibration mode
.
Dimensions 180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight 290 g (10 oz.)
REQUIRED REAGENTS
Code Description Quantity/test
HI 95747-0 Copper Low Range Reagent 1 packet
SPECIFICATIONS
Precision is how closely repeated measurements
agree with each other. Precision is usually expressed
as standard deviation (SD).
Accuracy is defined as the nearness of a test result to
the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure explains
these definitions.
In a laboratory using a standard solution of 0.750
mg/L Copper and a representative lot of reagent, an
operator obtained with a single instrument a standard
deviation of 0.015 mg/L.
PRECISION AND ACCURACY