INTERFERENCES
• Positive error: - Bromine
- Chlorine dioxide
- Iodine
- Ozone
- Oxidized Manganese and Chromium
Alkalinity above 250 mg/L CaCO
3
or acidity above 150 mg/L CaCO
3
will not reliably develop
the full amount of color or it may rapidly fade.
To resolve this, neutralize the sample with diluted HCl or NaOH.
In case of water with hardness greater than 500 mg/L CaCO
3
, shake the sample for
approximately 2 minutes after adding the powder reagent.
Chlorine
For free chlorine measurement
: add 3 drops
of HI 93701A-T DPD1 indicator and 3
drops of HI 93701B-T DPD1 buffer to an
empty cuvet.
Immediately add 10 mL of unreacted
sample. Replace the cap and shake gently.
Immediately reinsert the cuvet into the
holder and ensure that the notch on the
cap is positioned securely into the groove.
Press READ and “SIP” will blink during
measurement.
The instrument directly displays
concentration in mg/L of free chlorine on
the Liquid Crystal Display.
Then,
for total chlorine measurement
, remove
the cuvet and add to the reacted sample 3
drops of HI 93701C-T DPD3 solution.
Replace the cap and swirl gently.
Reinsert the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Wait for 2’ and 30” then press READ and
“SIP” will blink during measurement.
The instrument directly displays
concentration in mg/L of total chlorine on
the Liquid Crystal Display.
Remove the cuvet.