Manual

18 19
Press ZERO/SET and “SIP” will blink on
the display.
After a few seconds the display will show
“-0.0-”. Now the meter is zeroed and
ready for measurement.
Add 5 drops of HI 93710-0 reagent. Replace
the cap and mix the solution.
Reinsert the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Remove the cuvet.
Press READ and “SIP” will blink during
measurement.
The instrument directly displays the pH
value on the the Liquid Crystal Display.
pH
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
Range 6.5 to 8.0 pH
Resolution 0.1 pH
Precision ±0.1 @ 7.2 pH
Method Adaptation of the Phenol Red method. The reaction with the reagent causes a
yellow to red tint in the sample.
REQUIRE REAGENTSREQUIRE REAGENTS
REQUIRE REAGENTSREQUIRE REAGENTS
REQUIRE REAGENTS
Code Description Quantity/test
HI 93710-0 pH Reagent 5 drops
REAGENT SETSREAGENT SETS
REAGENT SETSREAGENT SETS
REAGENT SETS
HI 93710-01 Reagents for 100 tests
HI 93710-03 Reagents for 300 tests
For other accessories see page 24.
pH
MEASUREMENTMEASUREMENT
MEASUREMENTMEASUREMENT
MEASUREMENT
Fill the cuvet with 10 mL of unreacted
sample, up to the mark, and replace the
cap.
For most accurate results, use of a class A
laboratory pipette is recommended.
Place the cuvet into the holder and ensure
that the notch on the cap is positioned
securely into the groove.
Select the parameter code for pH “PH” by
holding ZERO/SET (see page 12).
pH