Instruction Manual HI 955201 • HI 955202 HI 955301 • HI 955302 1 or 2-Channel Printing and Logging Pt-100 Thermometers www.hannainst.
Dear Customer, Thank you for choosing a HANNA instruments® product. Please read this instruction manual carefully before using the instrument. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications. If you need additional technical support, do not hesitate to e-mail us at tech@hannainst.com These instruments are in compliance with the directives. TABLE OF CONTENTS Warranty ..........
GENERAL DESCRIPTION PRELIMINARY EXAMINATION For printing and measuring temperature in the higher ranges without having to compromise high resolution, HANNA instruments® has developed the HI 955201 single channel and the HI 955202 dual channel, 4-wire Pt100 printing thermometers. These thermometers have a resolution of 0.1°C in the range of -200.0 to 850.0°C.
FUNCTIONAL DESCRIPTION HI 955201 AND HI 955202 1 Range Resolution Accuracy 2 3 4 °C CH HI 955202 5 PT100 - PRINTING THERMOMETER DUAL CHANNEL 6 8 9 ON / OFF PAPER 10 7 INTV UP DOWN TEMP PRINT RECORD SPECIFICATIONS HI 955201 AND HI 955202 11 HI 955201 HI 955202 -200.0 to 850.0°C 0.1°C or 1°C, selectable ±0.1°C ±0.1% of reading (for 1 year, excluding probe error) Typical EMC Deviation ±0.
FUNCTIONAL DESCRIPTION HI 955301 AND HI 955302 Range Resolution Accuracy 1 2 3 4 °C CH HI 955302 5 PT100 - LOGGING THERMOMETER DUAL CHANNEL 8 6 9 PRINT PAPER ALT UP DOWN TEMP TIME LOG 10 7 1. 2. 3. 4. 5. 11 Connector(s) for Pt100 probe Power adapter plug Liquid Crystal Display PAPER key, to feed paper PRINT key, to print on-demand (current time and temperatures) 6. ALT key, to activate alternate function key 7.
INITIAL PREPARATION Each meter is supplied complete with batteries. Remove the back cover, unwrap the batteries and install them while paying attention to the correct polarity (see "Battery Replacement" section). To save battery life, the display is automatically switched off after 5 minutes of non-use. However, if in logging/recording mode, the meter will continue to monitor the temperature. To activate the display, press the ON/OFF (HI 955201 and HI 955202) or TEMP (HI 955301 and HI 955302) key.
SETTING MEASUREMENT RESOLUTION Press the TIME key and the day will be flashing. Press the TEMP key to select the 0.1°C or 1°C resolution. DATE TIME °C Set the correct day with the UP or DOWN key. CH TEMP With dual channel models, the resolution of both channel readings automatically change by pressing this key.
The time, date and printing interval are now set and stored in the memory even when the display is switched off. TO VIEW DATE, TIME & TEMPERATURE (for HI 955301 and HI 955302 only) To view the time, press the TIME key. The selected interval is also displayed on the secondary LCD. TIME INTV TIME To view the date, press the UP key while the LCD is displaying the time. DATE PRINTING / RECORDING WITH HI 955201 AND HI 955202 To print the measured values, press PRINT.
Data from the second channel can only be added if the recording mode is exited and a new log number is entered. A blinking "CH" will appear on the display next to the channel number not utilized and/or recorded. °C LOG CH When the meter is in recording mode, the "LOG" indication is displayed on the secondary display. To save battery life, if no key is pressed, the meter goes to stand-by mode. °C LOG CH To reactivate the display press ON/OFF.
To continue logging without printing, press the ALT and PAPER keys at the same time, and "LOG" will start blinking. °C PRINT PAPER SAMPLE NUMBER During logging it is possible to know the running sample number. Press LOG twice and the display will show the number of values stored in the current log. LOG LOG LOG ALT CH UP After approximately 5 minutes of inactivity, the display will switch off but the logging function remains active. To reactivate the display TEMP press TEMP key.
It is always possible to switch from the logging with printing function to the logging only function. Press the ALT and PAPER keys at the same time and the "LOG" symbol will start blinking to indicate that data are now only stored into memory. TO SCAN STORED DATA ON DISPLAY Press LOG and the display will show the log and page numbers of the next logging.
Press the DOWN key to revert back to sampling time and move back through samples. To exit the recall mode, press LOG. Note: This mode will not alter data into memory TO PRINT STORED DATA Once a log number is selected by using the ALT and DOWN keys, all or part of that log section can be printed by pressing ALT and PRINT simultaneously. The printer will print the logged section, beginning with the selected sample number and without altering the memory content.
The meter is now in printing/logging mode using log #1, page 16. The "LOG" and P 15 indications on the secondary display mean that the meter has entered the log mode and has 15 pages of available memory remaining. After the initial printout, the time and printing/logging interval will reappear. TIME INTV ALT UP DOWN TEMP TIME LOG ==01== 0010 10-01-96 0002M *00.18 Step 5 - Viewing data stored in sample #1 Press the LOG key to enter the memory recall mode.
Press UP again to view the temperature. DATA TRANSFER TO PC °C CH UP With HI 955302 when both probes are connected, press the UP key to view the other channel value. Step 6 - Printing stored data (samples #5 through #10) Press and hold the ALT key, then press DOWN until the total number of samples recorded in log #1 is displayed. ALT HI 955301 and HI 955302 feature infrared emitting circuitry.
HI 92000 allows you to use the powerful means of the most common spread sheet programs (e.g. Excel©, Lotus 1-2-3©). Simply run your favorite spread sheet and open the file downloaded by HI 92000. It will then possible to make any elaboration available with your software (e.g. graphics, statistic analysis). To install HI 92000, you need a 3.5" drive and a few minutes to follow the instructions conveniently printed on the disk's label.
PRINTER MAINTENANCE Allow approximately 5 cm (2") to exit from the printer and replace the cover. CHANGING THE INK CARTRIDGE When printouts become faint, it might be necessary to change the ink cartridge. Contact the nearest HANNA service center. INSERTING THE PAPER ROLL HI955201, HI955202, HI955301 and HI955302 use plain paper rolls, 38 mm width. To insert a new roll open the paper cover by pulling it gently. TO RESET PRINTER To reset a jammed printer, press the PAPER key.
BATTERY REPLACEMENT The "LO BAT" indication is displayed to warn the user that the batteries are running down. °C CH LOBAT If the warning appears during printing, it means that 200 printouts can still be made. When the battery level can only support 100 printouts, "LO BAT" is continuously displayed. Battery replacement must only take place in a safe area and using the battery type specified in this instruction manual.
CE DECLARATION OF CONFORMITY ACCESSORIES 4-WIRE PT100 PROBES HI 768A Air/gas probe, 1 m (3.3') cable HI 768L General purpose/liquid probe with 1 m (3.3') cable HI 768P Penetration probe, 1 m cable 205mm 8.1" 100mm 3.9" 3mm 0.12" HI 768P 205mm 8.1" 100mm 3.9" 3 6mm mm 0.24" 0.12" HI 768A 205mm 8.1" 100mm 3.9" 3mm 0.
SALES & TECHNICAL SERVICE Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 China: Tel. (10) 88570068 • Fax (10) 88570060 Egypt: Tel. & Fax (02) 2758.683 Germany: Tel. (07851) 9129-0 • Fax (07851) 9129-99 Greece: Tel. (210) 823.5192 • Fax (210) 884.0210 Indonesia: Tel. (21) 4584.2941 • Fax (21) 4584.2942 Japan: Tel. (03) 3258.9565 • Fax (03) 3258.9567 Korea: Tel. (02) 2278.5147 • Fax (02) 2264.1729 Malaysia: Tel. (603) 5638.9940 • Fax (603) 5638.9829 Singapore: Tel. 6296.7118 • Fax 6291.