User Guide
8
It is recommended to calibrate the instrument frequently, especially if
high accuracy is required.
The instrument should be recalibrated for pH:
a) Whenever the pH electrode or temperature probe is replaced.
b) At least once a month.
c) After testing aggressive chemicals.
d) If greatest accuracy is required.
e) Whenever the batteries have been replaced.
PREPARATION
Pour small quantities of pH 7.01 and pH 4.01 solution into two clean
beakers.
For accurate calibration use two beakers for each buffer solution, the
first one for rinsing the electrode and the second one for calibration.
This way the contamination of the buffers is minimized.
To obtain accurate readings, use pH 7.01 and pH 4.01 buffers if you
are going to measure acidic samples, or pH 7.01 and pH 10.01 for
alkaline measurements.
PROCEDURE:
In order to calibrate the meter, there is a choice of 5 memorized
buffers: 4.01, 6.86, 7.01, 9.18 and 10.01 pH.
• Make sure that the meter is in the pH mode.
• Remove the protective cap and rinse the electrode with some of
the buffer calibration solution that you are going to use first.
ONE-POINT CALIBRATION:
It is always recommended to perform a two-point calibration for
optimum accuracy. In case of necessity, it is possible to carry out only
a one-point calibration with the instruments. The offset buffers, that is
HI 7004 HI 7007
HI 7004
HI 7007
CALIBRATION
HI 7007
RINSE
pp
pp
p
H CALIBRATIONH CALIBRATION
H CALIBRATIONH CALIBRATION
H CALIBRATION










