User Guide
10
• Press the CAL key and the calibration process is ended with only
the offset of the meter calibrated. For best accuracy however, do
not press CAL and proceed with a two-point calibration.
TWO-POINT CALIBRATION
After calibrating at 6.86 or 7.01 pH (see above), for the second buffer
use pH 4.01 if you are going to measure acidic samples, and pH 9.18
or pH 10.01 if you are going to measure alkaline solutions.
• Proceed as described in "One-point calibration" above but do not
end the calibration by pressing CAL.
• After the first calibration point is confirmed, immerse the pH
electrode into the second buffer solution (pH 4.01, 9.18 or 10.01)
and stir gently.
Note: the electrode should be submerged approximately 4 cm (1½")
into the solution with the temperature probe located as close to
it as possible.
• Select the second buffer value on the secondary display by press-
ing the
Ï
°C or Ð
°C°C
°C°C
°C key.
• When the “READY” and “CON” symbols blink on the display, the
reading is stable and the calibration can be confirmed.
• Press the CFM key. If the reading is not close to the selected buffer
solution, “WRONG
2
BUF
” and “WRONG ” will blink alternatively.
If the reading is close to the selected buffer, the slope and the
offset are calibrated. The values will be stored in memory and the
meter will return to the operating mode.
Note: The meter will automatically skip the buffer that was used for
the first calibration to avoid erroneous calibration. At least 1
full pH unit is required between the two buffers used for the
offset and slope calibration. As a result, once calibrated at
either pH 6.86 or 7.01, the microprocessor will automatically
ignore the other one for the second point calibration and it will
do likewise for pH 9.18 and 10.01
Note: During calibration, the secondary LCD displays the selected
buffer value.










